background image

Features

Patented Thru-Coat™ technology  

 

measures and displays the thickness of 
the metal part and its coating using a 
‘single backwall echo’.

 Temperature Compensation feature 

 

adjusts the material velocity for changes 
in material temperature.

Optional Oxide/Scale feature  

 

measures and displays the thickness of 
the steel and the oxide/scale build-up.

Average/Min mode saves the average or 

 

minimum of several successive thick-
ness measurements.

A-scan display is brighter with better 

 

contrast and visibility.

Uses a vast selection of dual element 

 

and single element contact, delay line, 
and immersion transducers.

EMAT transducers for no-couplant steel 

 

thickness measurements through heavily 
scaled surfaces.

The file-based alphanumeric datalogger 

 

can use longer file names (32 character 
max) and ID numbers (20 character 
max).

Every thickness reading in a stored 

 

B-scan can now be reviewed in the 
GageView Interface Program and on the 
gage.

Select to view (Min/Max, Alarm, or A-

 

scan flags) flags for Grid points.

Grid files can be expanded by adding 

 

rows or columns or by changing the 
incrementing direction.

930-168-EN

37DL PLUS

Corrosion Thickness Gage

Advanced Ultrasonic Corrosion Gage

The Panametrics-NDT™ 37DL PLUS is 
an advanced nondestructive ultrasonic 
thickness gage that combines powerful 
measurement features with sophisticated 
data acquisition and output capabilities for 
applications involving pipes, tanks, and 
other metal structures subject to internal 
corrosion or erosion. The 37DL PLUS 
includes many innovative features that 
simplify true metal thickness measurements 
even when the exposed surface is coated 
or painted. The patented Thru-Coat 
technology measures and displays the 
thickness of the metal part and its coating 
using a ‘single backwall echo’. A Oxide/
Scale option measures and displays the 
thickness of the steel and the oxide/scale 
build-up on the inside of boiler tubes to 
help predict tube life. Accuracy of high 
temperature thickness measurements can 
be improved by using the Temperature 
Compensation feature that adjusts the 

material velocity for changes in material 
temperature. A new Average/Min mode 
saves the average or minimum of several 
successive thickness measurements. 
The A-scan display is bright with great 
contrast and visibility. Every thickness 
reading in a stored B-scan can be reviewed 
on the 37DL PLUS and in the Interface 
Program. The user can select to view (Min/
Max, Alarm, or A-scan flags) flags with 
Grid points. The file-based alphanumeric 
datalogger can use longer file names 
(32 character max) and ID numbers (20 
character max). Grid files can be expanded 
by adding rows or columns or by changing 
the incrementing direction.

Содержание 37DL PLUS

Страница 1: ...uctive ultrasonic thickness gage that combines powerful measurement features with sophisticated data acquisition and output capabilities for applications involving pipes tanks and other metal structures subject to internal corrosion or erosion The 37DL PLUS includes many innovative features that simplify true metal thickness measurements even when the exposed surface is coated or painted The paten...

Страница 2: ...ed Rt Angle 283 0 028 1 5 7 2 mm 0 71 37 mm M2017 20 LCM 74 4 Rt Angle 0 25 Steel 0 020 0 500 32 122 F 2127 6 35 mm 0 5 12 0 mm 0 50 C Oxide 0 010 0 050 0 25 1 25 mm M2091 20 LCM 74 4 Rt Angle 0 25 Steel 0 020 0 500 32 122 F 2127 6 35 mm 0 5 12 0 mm 0 50 C Oxide 0 006 0 050 0 15 1 25 mm E110 SB LCB 74 4 and Straight 1 25 0 080 5 00 32 176 F 1 2XA E110 28 5 mm 2 0 125 mm 0 80 C Dependent on materia...

Страница 3: ...boiler tube and the thickness of the oxide layer Knowing the thickness of the oxide scale helps predict tube life We recommend the use of the M2017 or M2091 transducer for this application Thru Coat Technology This innovative feature uses a single backwall echo to measure true metal thickness The 37DL PLUS can display both metal and coating thickness each adjusted for their correct material sound ...

Страница 4: ...2 1 lbs 0 95 Kg Internal Datalogger Datalogger and RS 232 The 37DL PLUS identifies stores recalls clears and transmits thickness readings waveform images and gage setup informa tion via the RS 232 serial port Baud Rate Word Length Stop Bits and Parity are adjustable from the keypad Maximum of Stored Values Standard 60 000 thickness readings or 4 500 waveforms with thickness readings File Names ID ...

Отзывы: