Olympus 37DL PLUS Скачать руководство пользователя страница 3

www.olympusNDT.com

Temperature Compensation

Variations in material temperature cause changes in sound velocity that may affect the 
accuracy of the thickness measurement. The Temperature Compensation feature allows 
the user to enter the calibration block temperature and also manually or automatically 
input the current (high) temperature at the measurement points. The 37DL PLUS will 
display the temperature corrected thickness that can then be saved to the internal 
datalogger.  

DB Grid View

The Grid View feature allows the user to select either a Min/Max, Alarm, or  
A-scan flag to be viewed along with each grid point. This allows the user to quickly scan 
through a file and locate Min/Max, Alarm, or A-scan locations.  

Optional Oxide/Scale Measurement

This option uses advanced algorithms to measure the thickness of oxide/scale build-up 
on the inside of boiler tubes. The gage simultaneously displays the metal thickness of 
the boiler tube and the thickness of the oxide layer. Knowing the thickness of the oxide/
scale helps predict tube life. We recommend the use of the M2017 or M2091 transducer 
for this application.

Thru-Coat™  Technology

This innovative feature uses a single backwall echo to measure true metal thickness. 
The 37DL PLUS can display both metal and coating thickness, each adjusted for their 
correct material sound velocity. The gage can be set to display only the true metal 
thickness. There is no need to remove the paint or coating from the surface. Thru-Coat 
measurements use the new D7906-SM and D7908 dual element transducers. 

The 37DL PLUS automatically calculates and displays 
both the coating thickness (0.019”) and the material 
thickness (0.284”) with a single backwall echo.

The thickness reading of 0.376” is compensated for the 
change in material sound velocity due to the tempera-
ture difference between the calibrated temperature and 
the material temperature (932°F).

Flags clearly indicate (H) High Alarm and (L) Low Alarm 
thickness measurements.

The 37DL PLUS displays the material thickness (0.164”) 
and scale thickness (0.011”) simultaneously.

Содержание 37DL PLUS

Страница 1: ...uctive ultrasonic thickness gage that combines powerful measurement features with sophisticated data acquisition and output capabilities for applications involving pipes tanks and other metal structures subject to internal corrosion or erosion The 37DL PLUS includes many innovative features that simplify true metal thickness measurements even when the exposed surface is coated or painted The paten...

Страница 2: ...ed Rt Angle 283 0 028 1 5 7 2 mm 0 71 37 mm M2017 20 LCM 74 4 Rt Angle 0 25 Steel 0 020 0 500 32 122 F 2127 6 35 mm 0 5 12 0 mm 0 50 C Oxide 0 010 0 050 0 25 1 25 mm M2091 20 LCM 74 4 Rt Angle 0 25 Steel 0 020 0 500 32 122 F 2127 6 35 mm 0 5 12 0 mm 0 50 C Oxide 0 006 0 050 0 15 1 25 mm E110 SB LCB 74 4 and Straight 1 25 0 080 5 00 32 176 F 1 2XA E110 28 5 mm 2 0 125 mm 0 80 C Dependent on materia...

Страница 3: ...boiler tube and the thickness of the oxide layer Knowing the thickness of the oxide scale helps predict tube life We recommend the use of the M2017 or M2091 transducer for this application Thru Coat Technology This innovative feature uses a single backwall echo to measure true metal thickness The 37DL PLUS can display both metal and coating thickness each adjusted for their correct material sound ...

Страница 4: ...2 1 lbs 0 95 Kg Internal Datalogger Datalogger and RS 232 The 37DL PLUS identifies stores recalls clears and transmits thickness readings waveform images and gage setup informa tion via the RS 232 serial port Baud Rate Word Length Stop Bits and Parity are adjustable from the keypad Maximum of Stored Values Standard 60 000 thickness readings or 4 500 waveforms with thickness readings File Names ID ...

Отзывы: