11 

 

The double bracket locks allow the coupling of 2 

accessories, taking up less space on the side rails. The 
plastic  handles  (1)  allow  height  and  rotation  on  the 
shaft  (distance  the  accessory  to  seat)  adjustment, 
while  the  metallic  lever  (3)  allows  the  bracket  lock 
adjustment on the side rails. 

Simply  turn  counterclockwise  to  loosen  and 

clockwise to lock. 

To adjust the bracket locks position on the side rail, pull the metal lever (3) counterclockwise. Adjust the 

attachment to the desired position and press the metal lever (3) clockwise. 

 

Note:  the  metal  lever  (3)  moves  only  in  the  area  indicated  in  the  figure  above.  To  turn  it  the  metal  lever  (3) 
clockwise,  push  it  upwards  causing  the  lever  to  pass  through  the  metallic  knob's  axis  (2),  and  then  rotate  it 
again. 
 

 

The Colposcope holder is attached to the metal support (4) of 

the  upholstered  leg  rest  or  knee  support,  optimizing  the 
professional's work space. This accessory can receive Colposcopes 
with pin fitting up to 22.5 mm diameter (7/8"). 

It has arm (3) with free joint, allowing better positioning of the 

Colposcope (5). 

To  attach  the  Colposcope:  fit  it  and  turn  the  knob  (1) 

clockwise by locking it in the desired position. 

To  adjust  the  height  (6):  turn  the  knob  (2)  to  loosen  the  regulatory  bar,  place  the  Colposcope  at  the 

requested height and turn the knob (2) clockwise to lock in position. 

 

 

They  offer  support  for  the  heels  and  are  available  with  double  bracket  locks  (3)  that  allow  the 

simultaneous fixation of the heels support (1) and the leg (

section 5.8

) rest or knees support (

section 5.14

). 

The heels support (2) can be adjusted for height, heel pads position and movement on the side rails. 
In  addition  to  bracket  locks  adjustments,  heel  supports  (2)  can  be  adjusted  for  the  heel  pad  distance 

according  to  the  patient  profile  or  procedure  to  be 
performed. 

To  fix  the  heel 

support  (2): 

fit 

the 

support  (2)  into  the   
holes (4) along the metal 
support  (1)  according  to 
the desired distance. 

 
 
 
 

Содержание VERSATILE

Страница 1: ......

Страница 2: ......

Страница 3: ...Operator and Others 8 5 5 Emergency Position 8 5 6 Paper Roll Holder Optional 9 5 7 Sliding Extender 9 5 8 Upholstered Leg Rest 9 5 9 Left and Right Bracket Locks 10 5 9 1 Accessories on Bracket Locks...

Страница 4: ...15 10 Technical Features 15 10 1 Electromagnetic Compatibility EMC 16 11 Dimensional 18 12 Symbology 19 13 Important Notes 19 13 1 General Cares Compulsory Reading 20 13 2 Disposal 21 13 3 Transport a...

Страница 5: ...ge Contents Check out the equipment package contents Standard Items 1 Versatile Table 1 Left Bracket Lock 1 Right Bracket Lock 2 Upholstered Leg Rest 1 Versatile Quick Operation Guide Optional Items 1...

Страница 6: ...wing items of the Versatile Table are considered as applied parts to the patient Standard Items Upholstery Upholstered ler rest Seat extension Extender upholstery Optional Items Auxiliary pillow Uphol...

Страница 7: ...lso it has a fuse holder for equipment 2 an auxiliary socket 3 and a fuse holder for this socket 4 To turn Off the equipment press the On Off switch 1 to the O position The headrest 1 was designed to...

Страница 8: ...0 centimeters remaining out of the equipment movement area and its components To ensure safe and proper patient positioning the operator should advise him to keep his hands in sight during the entire...

Страница 9: ...sliding extender 2 until its structure engages on the latches 3 To use the stainless tray 4 remove the extender upholstery 1 To use the upholstery 1 position it on the latches 3 To retract the sliding...

Страница 10: ...sh it upwards causing the lever to pass through the metallic knob s axis 5 and then rotate it again 5 9 1 Accessories on Bracket Locks Accessories which can be installed on the Versatile Table by usin...

Страница 11: ...optimizing the professional s work space This accessory can receive Colposcopes with pin fitting up to 22 5 mm diameter 7 8 It has arm 3 with free joint allowing better positioning of the Colposcope...

Страница 12: ...red with non slip coating it assists in the patient s boarding and allows the professional to better approach the equipment since the step 1 can be repositioned inside the cabinet For the professional...

Страница 13: ...tion systems that act on the mechanical and electrical parts of the set Pressurized suspension the high pressure nitrogen cylinder relieves the load of the gear motor lifting system of the equipment i...

Страница 14: ...proper functioning of the equipment and its components read Chapter 5 Description and Operating of the Equipment of this manual Installation should only be made by an authorized technician Installatio...

Страница 15: ...Power 118 127 V 200 VA Power 220 230 V 250 VA Protection fuses For 220 230 VC 1 5 A H 5 x 20 mm For 118 127 VC 3 A H 5 x 20 mm Power cable specification in compliance with requirements 6 1 and 6 2 of...

Страница 16: ...nment IMMUNITY Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wo...

Страница 17: ...environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measure filed strength in the location in which the Versatile Table is used exceeds the applicable...

Страница 18: ...18 11 Dimensional...

Страница 19: ...its publication Future technical improvements do not result in any right to update existing products The images presented in this manual are illustrative This equipment is designed to be free from in...

Страница 20: ...s not suitable for use in the presence of a flammable anesthetic mixture with air O2 or Nitrous Oxide This equipment should only be operated by physicians and nurses for clinical outpatient and gyneco...

Страница 21: ...s for transport and storage 20 to 70 This equipment is contraindicated for any use other than that for which it is intended or to be operated by unauthorized personnel 14 Troubleshooting For solving p...

Страница 22: ...ck the left and right bracket locks OPTIONAL ITEMS Check Colposcope holder joints Check the double bracket locks Check the armrests Check the knees support bracket lock Check heels support bracket loc...

Страница 23: ...Connection to wrong voltage power supply 5 This warranty will cease when a In the normal course of its period of warranty b Make changes in the product not authorized by Olsen c Adulterations in the p...

Страница 24: ...nd of low maintenance cost These characteristics have been achieved thanks to our competent and dedicated team which make me very proud for many reasons as they are always giving the best of their cre...

Страница 25: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Страница 26: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Страница 27: ......

Страница 28: ...55 48 2106 6000 export3 olsen odo br Registration at Ministry of Health of Brazilian 10281300011 Techinician In Charge Eng Cleber da Costa CREA SC S1 116283 5 Cod 5400400 Rev 05 17 05 2019...

Страница 29: ...VERSATILE...

Страница 30: ......

Страница 31: ...8 5 5 Posici n de Emergencia 8 5 6 Soporte para Rollo de Papel Desechable Opcional 9 5 7 Extensi n Deslizante 9 5 8 Pierneras Tapizadas 9 5 9 Abrazaderas Izquierda y Derecha 10 5 9 1 Accesorios sobre...

Страница 32: ...cnicas 15 10 1 Compatibilidad Electromagn tica EMC 16 11 Dimensional 18 12 Simbolog a 19 13 Notas Importantes 19 13 1 Cuidados Generales Lectura Obligatoria 20 13 2 Descarte 21 13 3 Transporte y Almac...

Страница 33: ...del Empaque A continuaci n observe el contenido del empaque Art culos Standard 1 Mesa Versatile 1 Abrazadera Izquierda 1 Abrazadera Derecha 2 Pierneras Tapizadas 1 Gu a R pida de Operaci n Mesa Versat...

Страница 34: ...oto 15 Cajones laterales 16 Abrazadera Derecha Se consideran partes aplicables al paciente los siguientes art culos Art culos de Serie Tapizados Pierneras tapizadas Opcionales Almohada auxiliar Apoyab...

Страница 35: ...ctrico tambi n est n situados el porta fusibles de protecci n del equipo 2 una toma auxiliar 3 y el porta fusibles de protecci n 4 de esta toma Para apagar el equipo presione el interruptor de On Off...

Страница 36: ...a del rea de movimiento tanto del equipo como de sus componentes La posici n segura y adecuada del paciente sobre el equipo la debe orientar el operador para que el paciente mantenga se con sus manos...

Страница 37: ...oclave de vapor Para acceder a estos elementos tire la extensi n deslizante 2 hasta que su estructura se encaje en las trancas Para utilizar la bandeja de acero inoxidable 4 retire el tapizado extenso...

Страница 38: ...arriba haciendo que la varilla met lica 4 pase por dentro de su eje 5 para luego girarla nuevamente 5 9 1 Accesorios sobre Abrazaderas Accesorios fijos en la Mesa Versatile por medio de abrazaderas C...

Страница 39: ...io para el trabajo del profesional Este accesorio puede recibir Colposcopios con pin de acople de hasta 22 5 mm de di metro 7 8 Presenta el brazo 3 con articulaci n libre lo que permite una mejor posi...

Страница 40: ...e en 24 horas la reducci n bacteriana es superior a 99 9 Carenados producidos en ABS de alta resistencia con cobertura en acr lico no necesitan de pintura lo que facilita el pulido superficial para re...

Страница 41: ...os para conectar el equipo a la red Si se han comprado accesorios opcionales estos tambi n deben instalarse en este momento El t cnico tambi n inspeccionar el equipo asegur ndose de que est de acuerdo...

Страница 42: ...sen no aconseja el uso de cualquier tipo de producto qu mico para la limpieza de estas partes sin embargo para desinfecci n es importante verificar si el producto presenta compatibilidad y caracter st...

Страница 43: ...cho 10 A 250 V NBR 14136 Tipo de protecci n contra choque el ctrico equipo Clase I de acuerdo con IEC 60601 1 y IEC 60601 1 2 Grado de protecci n Tipo B Modo de operaci n no continuo Tiempo On 30 s Ti...

Страница 44: ...ticas IEC 61000 4 2 6 kV por contacto 8 kV por aire 6 kV por contacto 8 kV por aire Los pisos deber an ser de madera u hormig n o estar provistos de baldosas de cer mica Si el suelo es de material si...

Страница 45: ...ico provocado por transmisores de RF fijos se deber a considerar la posibilidad de realizar una prueba electromagn tica in situ Si la intensidad de campo medida en el lugar donde habr de usarse la Mes...

Страница 46: ...18 11 Dimensional...

Страница 47: ...la fecha de su publicaci n Los perfeccionamientos t cnicos futuros no resultan en ning n derecho de cambios a productos ya existentes Las im genes presentadas en este manual son para fines ilustrativ...

Страница 48: ...t sica inflamable con aire O2 u xido Nitroso Tampoco es adecuado para uso en el entorno rico en Ox geno Este equipo se debe operar solamente por m dicos y enfermeros para ex menes cl nicos ambulatorio...

Страница 49: ...e y almacenamiento 00 a 45 C No mueva y no guarde el equipo en superficies irregulares Este equipo est contraindicado para cualquier uso distinto de aquel para el que se pretende o para ser operado po...

Страница 50: ...ones Verificaci n de las abrazaderas izquierda y derecha ART CULOS OPCIONALES Verificaci n de las articulaciones del soporte de colposcopio Verificaci n de las abrazaderas dobles Verificaci n de los a...

Страница 51: ...n a la red el ctrica de energ a con tensi n incorrecta 5 Esta garant a cesar a Por el decurso normal de su plazo de validez b Por alteraciones no autorizadas por Olsen en el producto c Por adulteraci...

Страница 52: ...de mantenimiento muy bajo Estas cualidades se alcanzar n a trav s de un equipo competente y dedicado del cual me enorgullezco sobre todos los aspectos dispuesto a llevar a nuestros clientes lo mejor...

Страница 53: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Страница 54: ...____________________________________ __________________________________________ __________________________________________ __________________________________________ __________________________________...

Страница 55: ......

Страница 56: ...55 48 2106 6000 export3 olsen odo br Registro en el Ministerio de Salud de Brasil 10281300011 Responsable T cnico Ing Cleber da Costa CREA SC S1 116283 5 C d 5400400 Rev 06 22 03 2021...

Отзывы: