Sensors (non-exhaustive list)
Table No. 1
Methane
CH4
Propane
C
3
H
8
Methane
CH4
Oxygen
O
2
2 years
Oxygen
O
2
1 year
Carbon Dioxide
CO
2
Carbon
Monoxide
CO
Hydrogen
sulphide H
2
S
Sensor reference
6 313 888
6 313 888
6 313 889
6 313 780
6 313 817
6 313 818
6 313 787
6 313 788
Standard range (1)
0 - 100%
LEL CH4
0 - 100%
LEL C
3
H
8
0 - 100% vol
2 - 30 % volume
2 – 30 % volume
0 - 5 % v/v
1000
100
Measurement principle
Thermo -
catalytic
Thermo -
catalytic
Catharo
metric
Electrochemical
Electrochemical
Infrared
absorption
Electrochemica
l
Electrochemica
l
Display resolution (1)
1 % LEL
1 % LEL
1% v/v
0,1 % v/v
0,1 % v/v
0,1 % v/v
1
1
Accuracy (2)
2
2
2
0.3 % v/v
0.3 % v/v
0.2 % v/v
15
3
Repeatability (3)
± 1 % LEL
± 1 % LEL
± 1 % vol
0,1 % v/v
0,1 % v/v
0.1 % v/v
1
1
Deviation of zero/sensitivity (4)
0,5 / 5
0,5 / 5
0,.2 / 2
0,2 / 2
0.2 / 2
0,2 / 2
0.5 / 1,5
0.5 / 2,5
Response time (5)
< 20 seconds
< 25 seconds
< 20 seconds
< 10
< 10
< 30
< 30
< 25
Temperature (6)
-20°C to
+50°C
-20°C to
+50°C
-20°C to +50°C
-20°C to +40°C
-20°C to +40°C
-10°C to40°C
-20°C to40°C
-20°C to40°C
Relative humidity and pressure
range (7)
0 – 95 % RH
1 bar ± 20 %
0 – 95 % RH
1 bar ± 20 %
0 – 95 % RH
1 bar ± 20 %
10 – 95 % RH
1 bar ± 20 %
10 – 95 % RH
1 bar ± 20 %
10 – 95 % RH
1 bar ± 20 %
10 – 95 % RH
1 bar ± 20 %
10 – 95 % RH
1 bar ± 20 %
Service life (8)
48 months
48 months
60 months
28
16
60
36
36
Storage conditions and maximum
storage time (9)
-40°C to +40
°C
10-60 % RH
1 bar ± 10 %
6 months
maximum
-40°C to +40
°C
10-60 % RH
1 bar ± 10 %
6 months
maximum
-40°C to +40
°C
10-60 % RH
1 bar ± 10 %
6 months
maximum
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
3 months
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
3 months
0 – 40 °C
10 – 60 % RH
1 bar ± 20 %
6 months
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
2 months
4 – 20 °C
10 – 60 % RH
1 bar ± 10 %
2 months
Response time (10)
30 s
30 s
Sensors functional immediately after start up of
device
120 s
Sensors functional immediately
after start up of device
Notes:
- Measurement is
underestimated if oxygen
level is
<
10%
- Exposure to high levels of
silicon or sulphur vapors
may damage the detector.
- The detector is sensitive to
a majority of explosive gases
- Presence of high levels of CO
2
can lead to an
over estimation of O
2
concentration
- Exposure to high levels of
organic solvents can damage the
sensors.
- Exposure to gases at higher
levels than detector’s range can
damage it Recalibrate sensors if
they go out of range.