Connect the
blue labeled end
of TUBE-A to LEO gas inlet port
labeled with icon .
Put the cap on LEO gas inlet port labeled with icon .
Connect the
red labeled end
of TUBE-A to the incubator or the point
where measurement is required.
Use the fitting tubes and/or additional tube in order to allow the
compatibility of the supplied tubing with the gas port of your
incubators.
If the incubator has the return input available, use the Output Tube
to connect the LEO gas outlet port labeled with icon to the
incubator gas recycle.
Press Pump/Diffusion Mode button to select the Pump
Mode.
Start a single point or logging gas measurement.
IN
1. CO2-O2 Measurement
ASPIRATION MODE
2
1
4
3
1
2
3
4
5
TUBE-A
IN
OUT
Connect the
blue labeled end
of TUBE-A to LEO gas inlet port labeled with icon .
Put the cap on LEO gas inlet port labeled with icon .
Connect the
red labeled end
of TUBE-A to the incubator or the point where measurement is required.
Use the fitting tubes and/or additional tube in order to allow the compatibility of the supplied tubing with the gas port
of your incubators.
Do not exceed 300 mbarg upstream to LEO.
Press Pump/Diffusion Mode button to select the Diffusion Mode, and start a single point or logging gas
measurement.
IN
DIFFUSION MODE
1
2
1
2
3
4
TUBE-A
IN
FOR HUMID GAS
TUBE-B
TUBE-A
If the sampling gas is humid, connect the
red labeled end
of TUBE-A and the
green connector
of TUBE-B to the Moisture Trap.
Connect the end of Tube-B to the incubator or to the point where measurement
is required.
Press Dry/Wet Mode button to set the Wet Mode,
and insert the humidity and temperature values of your
incubator.
Before starting a Gas Logging, unscrew the condensation trap to make sure the
condensation trap top lid has been removed.
The condensation trap top lid
allows reducing the sampling gas
time during a Gas Single Point
measurement Gas Measurement
with Moisture Trap.
2
3
1
Always take care
of the room
ventilation.
6
Use the fitting tubes and/or additional tube in order to allow the compatibility of
the supplied tubing with the gas port of your incubators.
Please verify the correct measurement mode for each incubator.