background image

3-3

6600-0371-000

10/27/03

3/Operation

3-3

This switch is used to select either the Manual or Servo mode of operation. An audio tone
sounds momentarily when the mode switch is depressed.

Increase (

M

M

M

M

M

) and Decrease (

?

?

?

?

?

) Switches:

These switches are used to set the radiant power level (% power) in the manual mode
and to set the control temperature in the servo mode.

Alarm Silence:

This switch is used to silence the audible alarm. It silences all alarms except the system
failure alarm. Refer to Table 1 - Warmer Alarms for details on how long the audible alarms
are silenced.

Holding this switch depressed for five seconds or longer activates the audible alarm and
lights all the indicator lamps for testing purposes.

Start/Hold:

This switch activates the elapsed timer. Refer to the elapsed timer operation, Section 3.3,
for a detailed explanation of its function.

Reset:

This switch is used to reset the elapsed timer to “00:00”.

Apgar Tones On/Off:

This switch activates the Apgar tones. When the Apgar tones are activated, a short alarm
tone is sounded at one minute and every five minute interval after the elapsed timer is
started.

Light On/Off:

This switch activates the observation light located in the heater assembly.

WARNING:

w

Prolonged exposure to the light emitted by the observation lamp in this
unit may harm the unprotected eyes of the infant. For safety, cover the
infant’s eyes.

Raise Bed (

M

M

M

M

M

(Elevating Models Only)

WARNING:

w

Check for proper clearance above the Warmer and below the bed surface
before raising or lowering the bed.

w

Inspect all patient connected tubes or wires before and after moving or
tilting the bed. Tilting or moving the warmer bed up or down can pull on
tubing or leads connected to the patient. This may disconnect tubes or
leads, restrict gas or liquid flow, or move sensors out of position.

CAUTION:

w

Do not continue to run the motor at the upper and lower limit positions; equip-
ment damage may result.

This switch raises the bed to a maximum height of 118 cm from the floor.

Lower Bed  (

?

?

?

?

?

(Elevating Models Only)

This switch lowers the bed to a minimum height of 98 cm from the floor.

6600−0371−000−DWG

100

MANUAL−DOC, User Reference Manual, OM Panda, English−default

Reproduced from the electronic master in MATRIX

Содержание Panda

Страница 1: ...mer OperationandMaintenanceManual 6600 0371 000 DWG 100 MANUAL DOC User Reference Manual OM Panda English default Reproduced from the electronic master in MATRIX Approved 2009 6 4 16 13 EET DST Barbar...

Страница 2: ...to the nearest Ohmeda Medical Regional Service Center This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Ohmeda Medical and by Ohme...

Страница 3: ...CheckoutProcedure 2 4 A Control Unit Check 2 4 B ElapsedTimerCheck 2 5 C ObservationLightCheck 2 5 D Raise and Lower Bed Switch Check 2 5 E Interlock Switch Check 2 6 F Power Failure Memory and Batter...

Страница 4: ...y 5 5 5 5 Battery Test 5 5 5 6 Maintenance Schedule 5 6 A Operator Maintenance 5 6 B Service Maintenance 5 6 6 Illustrated Parts Appendix Temperature Conversion Chart A 1 Servo Mode Algorithm A 2 Infa...

Страница 5: ...is used when the possibility of injury to the patient or the operator exists Symbols Attention Consult the manual for more information Type B equipment FunctionalEarthTerminal ProtectionEarthTerminal...

Страница 6: ...iv 6600 0371 000 10 27 03 Table of Contents iv Notes 6600 0371 000 DWG 100 MANUAL DOC User Reference Manual OM Panda English default Reproduced from the electronic master in MATRIX...

Страница 7: ...rocedures Mechanical and Control Unit while a patient occupies the Warmer Complete the Checkout Procedures section of this manual before putting the unit into operation If the warmer fails any portion...

Страница 8: ...vi 6600 0371 000 10 27 03 Table of Contents vi Notes 6600 0371 000 DWG 100 MANUAL DOC User Reference Manual OM Panda English default Reproduced from the electronic master in MATRIX...

Страница 9: ...accessory items 1 3 Heater Assembly Thewarmerisdesignedtowarminfantsonabedsurfacelocatedbeneaththeheaterassembly WARNING w Bed to heater spacing less than the recommended 64 cm will result in incor r...

Страница 10: ...le bed platformallowsTrendelenburgandreverseTrendelenburgpositioning Thehydraulic system for the tiltable bed provides a smooth dampened motion to avoid disturbing the patient 1 6 Bassinet Two bassine...

Страница 11: ...rocedure WARNING w Before using the Ohmeda Panda Baby Warmer read this entire manual Attempting to use this device without a thorough understanding of its operation may result in patient or user injur...

Страница 12: ...back to the normal position Check for a smooth rotation C Mechanical Checks 1 Check the operation of the bed sides The bed sides should operate smoothly WARNING w Regularly inspect the bed side panel...

Страница 13: ...ator Alarm Lights Alarm Silence Mode Switch Mode Indicators Bed Controls Bed controls present on elevating bed models only CI 14 096 Figure 2 1 Wall mount pin insertion 1 Verify that the warmer is rig...

Страница 14: ...prompt tones sound and the power display flashes 2 Adjust the heat output with the increase M and decrease touch switches to attain the high and low limits as indicated by the power display 3 Connect...

Страница 15: ...ch switch to activate the elapsed timer Verify that the timer starts operating 2 Press the Apgar tones On Off touch switch for the Apgar tones Verify that the indicator light for the Apgar tones is ex...

Страница 16: ...10 minutes test or partially charged to pass the two minute test If the battery is defective replace it Refer to the service manual There is no maintenance required for the battery The battery has a...

Страница 17: ...erature sensed at the skin tempera ture probe tip The skin temperature probe must be properly connected to the warmer and the infant for an accurate patient temperature measurement The range of temper...

Страница 18: ...st take the responsibility for making heater adjustments or environmental changes in response to the following conditions patient size and weight bed to heater head height differences changes in the e...

Страница 19: ...alarm tone is sounded at one minute and every five minute interval after the elapsed timer is started LightOn Off This switch activates the observation light located in the heater assembly WARNING w...

Страница 20: ...owerFailureAlarm Probe Failure Alarm The probe failure alarm is only active in the servo mode of operation The alarm activates when the skin temperature probe fails electrically due to an open or shor...

Отзывы: