3-1
6600-0371-000
10/27/03
3/Operation
3-1
In this section
3.1
Control Panel Operation ....................................................................................... 3-1
A. Displays ......................................................................................................... 3-1
B. Indicator Lights ............................................................................................... 3-2
C. Switches ........................................................................................................ 3-2
D. Alarms ............................................................................................................ 3-4
3.2
Preheat, Manual and Servo Mode Operation
..................................................... 3-7
3.3
Elapsed Timer Operation .................................................................................... 3-11
3.4
Bed Platform Operation ...................................................................................... 3-11
3.5
Side Panel Operation ......................................................................................... 3-12
3.6
Oxygen Administration ....................................................................................... 3-13
3.7
Gas Cylinder Installation and Operation ............................................................. 3-14
3.8
Mounting Accessories ....................................................................................... 3-16
A. Mounting and Releasing Procedure .............................................................. 3-17
B. Adapter Plate Mounting and Releasing Procedure ........................................ 3-17
3.9
Rotating Drawer Option ...................................................................................... 3-18
3.1 Control Panel Operation
(Control Panel is shown in Figure 2-2)
A. Displays
• Patient Temperature
• Control Temperature*
• % Power
• Elapsed Time*
*
During the preheat mode "PrE HEAT" flashes in the control temperature and elapsed time displays.
Patient Temperature:
The patient temperature display indicates the temperature sensed at the skin tempera-
ture probe tip. The skin temperature probe must be properly connected to the warmer and
the infant for an accurate patient temperature measurement. The range of temperature
measurement is from 30.0 to 42.0°C, with a resolution of 0.1°C. Temperatures above this
range result in an “HH.H” patient temperature display and below this range result in an
“LL.L” patient temperature display .
Control Temperature:
The control temperature display indicates the control temperature you select in the servo
mode. The servo control temperature range is from 35.0 to 37.5°C. The control tempera-
ture display is not active in the manual mode.
% Power:
The % power display indicates the percentage of maximum power that is being supplied
to the radiant heater in 5% increments.
Power settings between 0 to 25% allow operation without the 12 minute check patient
alarm in the manual mode.
6600−0371−000−DWG
100
MANUAL−DOC, User Reference Manual, OM Panda, English−default