85
Appendix A - special applications
Enter
The above illustration represents 3 sieves which weighed between 3.4 and
3.9 grams each. TARE SETUP was set OFF, AUTO SAMPLE was set OFF and
% RETENTION was selected. The sample weight retained in each different
sieve weighed between 3.839 and 4.377 grams. The total amount of
retention is shown in percent on the last display which was 0.623 for three
sieves. Initial sample was 200.25 grams.
N/A
NA
3.426 3.839
N/A
N/A
3.969 4.388
N/A
N/A
3.962 4.377
CONTINUE SAVE RESAMPLE DELETE
TAREWT INIT WT TARE WT FINAL \VT
SAMPLE #1
With CONTINUE highlighted, press
Enter
button to review final display.
PRESS <ENTER> FOR MENU
SAMPLE # INIT WT FINAL WT % RETENSION
1
200.025 0.413 0.206
2
200.025 0.419 0.209
3
200.025 0.415 0.207
TOTAL
1.247 0.623
Final display indicates the initial weight, final weight of all sieves and the
total weight and total amount of retention in percent.
Final weighing of sieves
After pressing
Enter
button, the display returns to final weighing. Place first
full sieve on the pan and press
Enter
button. Repeat weighing with remaining
sieves pressing the
Enter
button after each sieve. When the last sieve has
been weighed, the display indicates the final weight of all samples.
3.841
DIFF WEIGHING
TARE WT:
TARE WT:
INIT WT: FINAL WT:3.841
PUT FULL SIEVE #1 ON PAN STABLE
START/SAMPLE EDIT SETUP
Enter
DIFF WEIGHING SETUP
LIBRARY NAME
SIEVE 2
TARE SETUP
OFF
AUTO SAMPLE
ON
DIFF RESULT
% RETENTION
NUMBER OF SAMPLES
3
VIEW RESULTS
CLEAR ALL DATA
RUN
3.426
DIFF WEIGHING
TARE WT:
TARE WT:3.426
IN IT WT:
FINAL WT:
PUT EMPTY SIEVE # 1 ON PAN STABLE
START/SAMPLE EDIT SETUP
Initial weighing of sieves
Press the
Enter
button, the display indicates to PUT EMPTY SIEVE #1 ON
PAN.
Place empty sieve #1 on the pan and press
Enter
button. When AUTO
SAMPLE is set ON, display advances to the next sieve. Continue adding
sieves until all sieves have been weighed pressing the
Enter
button after
each sieve. When the last sieve has been weighed, the display indicates the
tare weight of all sieves. Press the
Enter
button.
This sample display indicates 3 sieves were weighed. With
CONTINUE
highlighted at the bottom of the display, press
Enter
button. The display
advances and indicates PUT FULL SIEVE #1 ON PAN.
N/A
NA
3.426 0.000
N/A
N/A
3.969 0.000
N/A
N/A
3.962 0.000
CONTINUE SAVE RESAMPLE DELETE
TAREWT INIT WT TARE WT FINAL \VT
SAMPLE #1
Enter