VN Series Floor Scale EN-3
OPERATION
For setup, calibration and proper operation, refer to the instruction manual supplied with the indicator. Refer to the specifications
section of the floor scale for capacity, and readability information.
For continued best operating performance:
•
Avoid falling loads, shocks and hard lateral impacts
•
When weighing, all items should be centrally located on the floor scale and not over
hanging the sides or the ramps.
•
Verify the calibration of the floor scale system at regular intervals.
Operating Limits
Although the floor scales have extremely rugged construction, certain load limits must not be exceeded. Depending on the type of
loading, the maximum safe static load, i.e. the maximum admissible loading, is shown in Table 2.
TABLE 2. MAXIMUM ADMISSIBLE LOADING.
Model
Maximum Central load
1
(80% Surface Area)
Maximum Side load
2
(30% Surface Area)
One Side Corner Load
3
(10% Surface Area)
VN5000L
8000 lb
5000 lb
2600 lb
VN5000X
8000 lb
5000 lb
2600 lb
SEALING METHODS
Figure 2. Sealing Method 1
Figure 3. Sealing Method 2
Содержание VN Series
Страница 2: ......
Страница 12: ...EN 10 VN Series Floor Scale...
Страница 22: ...ES 10 B scula para piso serie VN...
Страница 32: ......