Valor 2000 V22PW FAQ
AppNotes.v20180720_ AG
For any further questions, please e-mail us at [email protected], or call us at 973-377-9000 #2 or ext. 7852 from 8AM to 5PM EST
Topic:
This list compiles frequently asked questions about the Valor 2000 V22PW
(plastic housing only) model:
a) How do I switch between units of measure
b) How do I select and display the active unit of measure
c) How do I calibrate my unit
d) How do I clear an Err 9 Data error?
How to switch between units of measure
With this model, you can only select one unit of measure at a time, which you need to activate from the
Service menu (see next section:
How to select and display the active unit of measure
to view a short video on this topic.
How to select/display the active unit of measure
Enabling or disabling the units of measure can be accomplished in the SERVICE > S.UNIT sub-menu
(you can also download the user manual
– refer to Section 3.5: Unit Menu).
to view a short video on this topic.
To enable another unit of measure, you need to disable the current active unit first (the unit displayed
when the scale is turned on).
For the following example, the current active unit is grams (g), and you want to change to pounds (lb):
1. With the scale off, press and hold [On/Zero] and [Tare]
together for up to 15 seconds
– the scale will cycle through
power up and weigh mode before RAMP is displayed.
2. Release the keys after RAMP is displayed.
3. When RAMP is displayed
– press [Tare] (no) repeatedly
through each succeeding sub-menus
*
(LIN > SPAN > CAP >
DIV5 > L.IR FL) until S.UNIT is displayed.
* DO NOT press [On/Zero]
as these sub-menus are displayed.
Notes:
a) If you overshoot the S.UNIT sub-menu, just keep pressing [Tare] until S.UNIT is displayed again.
b) To abort at any time, long press [On/Zero] to turn the scale off. Repeat from Step 1.