Valor 2000 V22PW FAQ
AppNotes.v20180720_ AG
For any further questions, please e-mail us at [email protected], or call us at 973-377-9000 #2 or ext. 7852 from 8AM to 5PM EST
6. Place the required calibration mass on the pan, then press
[On/Zero].
The display will flash
–C-- while the reading is stored, then
shows DONE when the calibration is completed.
7. The scale will display the next sub-menu: CAP
8. When CAP is displayed
– press [Tare] repeatedly through
each succeeding sub-menus (DIV5 > L.IR FL > S.UNIT >
E.PAND > RESET > END) until END is displayed*.
* If you overshoot END, just keep pressing [Tare] until END is
displayed again.
9. When END is displayed
– press [On/Zero] to exit the service
menu.
10. The scale returns to weighing mode and the error is cleared.
If this still does not resolve this issue, please contact Ohaus Technical Support.