Valor 2000 V22PW FAQ 

AppNotes.v20180720_ AG 

For any further questions, please e-mail us at [email protected], or call us at 973-377-9000 #2 or ext. 7852 from 8AM to 5PM EST 

 

6.  Place the required calibration mass on the pan, then press 

[On/Zero]. 
 

The display will flash 

–C-- while the reading is stored, then 

shows DONE when the calibration is completed. 

 

 

7.  The scale will display the next sub-menu: CAP 

 

8.  When CAP is displayed 

– press [Tare] repeatedly through 

each succeeding sub-menus (DIV5 > L.IR FL > S.UNIT > 
E.PAND > RESET > END) until END is displayed*. 

 

* If you overshoot END, just keep pressing [Tare] until END is 
displayed again. 

 

 

9.  When END is displayed 

– press [On/Zero] to exit the service 

menu. 

 

 

10.  The scale returns to weighing mode and the error is cleared. 

 

 

 

If this still does not resolve this issue, please contact Ohaus Technical Support.

 