Valor 2000 V22PW FAQ
AppNotes.v20180720_ AG
For any further questions, please e-mail us at [email protected], or call us at 973-377-9000 #2 or ext. 7852 from 8AM to 5PM EST
How to clear Err 9 Data error (service span cal)
An Err 9 data error can occur if the internal service settings were somehow altered. This can be cleared
with a service span calibration.
Before starting calibration, be sure you have the appropriate calibration masses. See below table for your
particular model:
Note: if the active unit of measure is lb or oz before initiating calibration, use lb calibration mass. If in g or kg,
use kg calibration mass.
For this example, we will use a V22PWE15T (15kg/30lb model):
1. With the scale off, press and hold [On/Zero] and [Tare]
together for up to 15 seconds
– the scale will cycle through
power up and weigh mode before RAMP is displayed.
2. Release the keys after RAMP is displayed.
3. When RAMP is displayed
– press [Tare] (no) until SPAN is
displayed.
Notes:
a) If you overshoot the SPAN sub-menu, just keep pressing [Tare] until SPAN is displayed again.
b) To abort at any time, long press [On/Zero] to turn the scale off. Repeat from Step 1.
4. When SPAN is displayed:
a) Make sure there is nothing on the pan, then
b) Press [On/Zero] (yes) to initiate calibration
The display will flash
–C-- until the zero reading is stored, then
shows the requested calibration value (30lb in this example)