How to clear Err 9 Data error (service span cal) 

An Err 9 data error can occur if the internal service settings were somehow altered.  This can be cleared 
with a service span calibration.   

Before starting calibration, be sure you have the appropriate calibration masses.  See below table for your 
particular model: 

 

Note: if the active unit of measure is lb or oz before initiating calibration, use lb calibration mass.  If in g or kg, 
use kg calibration mass.   

 

For this example, we will use a V22PWE15T (15kg/30lb model): 

1.  With the scale off, press and hold [On/Zero] and [Tare] 

together for up to 15 seconds 

– the scale will cycle through 

power up and weigh mode before RAMP is displayed. 

 

 

 

2.  Release the keys after RAMP is displayed. 

 

3.  When RAMP is displayed 

– press [Tare] (no) until SPAN is 

displayed. 

 

Notes: 

a)  If you overshoot the SPAN sub-menu, just keep pressing [Tare] until SPAN is displayed again. 
b)  To abort at any time, long press [On/Zero] to turn the scale off.  Repeat from Step 1. 

 

4.  When SPAN is displayed: 

a)  Make sure there is nothing on the pan, then 
b)  Press [On/Zero] (yes) to initiate calibration 

 

The display will flash 

–C-- until the zero reading is stored, then 

shows the requested calibration value (30lb in this example) 

 

 

 

 

 