APPENDIX A STANDARD CALIBRATION
Ohaus Corporation www.ohaus.com
A-2
Valor
™ 1000W Series Service Manual
2d). Press
ZERO
to accept the setting shown, or
M+
or
TARE
to move
up or down to match value for specific model
– see table above.
6. Press
ZERO
to display
Ex
(x=Checkweigh mode; 0 is off; x=1 is on).
Press
ZERO
to accept
0
, or
M+
or
TARE
switch,
then press
ZERO
to accept.
E___3
7. The display will show
Load 0
. Make sure pan is empty,
then press
ZERO
to accept.
CAL 0
appears, then
Load F
.
Load 0
*
If span weight is not available, select a lower value: press
Select
, then
TARE
to choose
desired value. Complete process in Step 8.
EndCAL
appears. Scale returns to weighing mode.
Note:
To change
only
the division setting (Step 2 in the calibration routine), press and hold
ZERO
and
SELECT
, then release
SELECT
before
ZERO
. Choose as in Step 2.
Содержание Valor 1000W