APPENDIX A STANDARD CALIBRATION
Valor
™ 1000W Series Service Manual
A-1
Ohaus Corporation www.ohaus.com
APPENDIX A.
STANDARD CALIBRATION & SETUP
A.1 SETUP AND CALIBRATION
Standard calibration should be performed prior to using a scale, and after service.
A 3kg mass is required to perform span calibration on all models.
NOTE:
Be careful not to touch the scale or the table while calibration is in progress,
as it will cause the process to fail.
Preliminary Steps
:
1. Be sure the scale is level and stable during the entire calibration process.
2. Allow the scale to warm up for approximately five minutes after stabilizing to room
temperature.
3. To abort calibration, turn the scale off anytime during the calibration process.
TABLE A-2. INITIAL CONFIGURATION SETTINGS FOR EACH MODEL
MODEL
Division
(A----x)
Decimal point
(b----x)
Capacity
(c-----x)
Zero tracking
(d----x)
Calibration
Weight
*
V11PW3
0.5g: x=5
Kg: x=4 or g: x=1
3.0000kg or 3000.0g
x=3
3.000
V11PW6
1g: x=1
Kg:x=3 or g: x=0
6.000kg or 6000g
x=3
3.000
V11PW15
2g: x=2
Kg: x=3 or g:
x=03333333
15.000kg or 15000g
x=3
3.000
V11PW30
5g: x=5
Kg: x=3 or g: x=0
30.000kg or 30000g
x=3
3.000
Note:
Be sure units are set to
kg
before starting calibration.
1. Turn on the scale. When display sequence flashes, press and hold
ZERO
and
SELECT
until (
999999
) appears. Then
CAL---
appears.
SET---
2. Press
ZERO
to display
A____x
(x=divisions in grams: 1,2,5,10,20,50).
Press
ZERO
to accept the setting shown, or press
M+
or
TARE
to
increment or decrement setting to match value for specific model
–
see table above.
Note:
Settings for each model must match table above.
A___2
To change
only
division setting, press and hold
ZERO
and
SELECT
at
power on< then release
SELECT
before
ZERO
. Choose setting as in
Step 2 above.
3. Press
ZERO
to display
b____x
(x=decimal point: 0,1,2,3).
Press
ZERO
to accept the setting shown, or press
M+
or
TARE
to
move up or down to match value for specific model
– see table above.
B___3
4. Press
ZERO
to display
Cxxxxx
(xxxxx= capacity of scale).
Press
ZERO
to accept the setting shown, or press
M+
or
TARE
to
move up or down to match value for specific model
– see table above.
C 15.000
5. Press
ZERO
to display
dx
(x=Zero tracking setting: 0-9 from 0.125d to
D___3
Содержание Valor 1000W
Страница 2: ......
Страница 4: ......
Страница 18: ...CHAPTER 2 DIAGNOSTIC GUIDE Ohaus Corporation www ohaus com 2 6 Valor 1000W Series Service Manual...
Страница 24: ...CHAPTER 3 MAINTENANCE PROCEDURES Ohaus Corporation www ohaus com 3 6 Valor 1000W Series Service Manual...
Страница 40: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 10 Valor 1000W Series Service Manual...
Страница 43: ......
Страница 44: ...P N XXXXXXXX SERVICE MANUAL Valor 1000W SERIES SCALES...