CHAPTER 1 GETTING STARTED
Ohaus Corporation www.ohaus.com
1-6
Valor
™ 1000W Series Service Manual
Figure 1-4. Route of electrical Load Cell input and output through Wheatstone Bridge circuit.
A Load Cell is a force sensor, which receives a voltage (excitation) from a regulated power
source in the Indicator, and sends back a low-voltage milli-volt (mV) relative to the force applied.
The Load Cell signal is read by the Indicator, which converts it to a numeric value and shows it
visually. This output value increases as weight is loaded to the Load Cell.
For example, a 500kg Load Cell with 2mV/V output and 5V excitation would have a change of
0.02mV per kg change in the load, that is, 10mV/500=0.02mV per kg. At a load of 250kg the
signal voltage would thus increase by 5mV from the value measured at no load.
Tension
Tension
Compression
Compression
+Excitation
(Supply)
–
Excitation
(Supply)
INPUT(volts)
LOADCELL
2 mV/V
Strain
Gauges
4 strain gauges
connected as a
Wheatstone
Bridge
OUTPUT (milli volts)
+
Sig
–
Sig
Содержание Valor 1000W
Страница 2: ......
Страница 4: ......
Страница 18: ...CHAPTER 2 DIAGNOSTIC GUIDE Ohaus Corporation www ohaus com 2 6 Valor 1000W Series Service Manual...
Страница 24: ...CHAPTER 3 MAINTENANCE PROCEDURES Ohaus Corporation www ohaus com 3 6 Valor 1000W Series Service Manual...
Страница 40: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 10 Valor 1000W Series Service Manual...
Страница 43: ......
Страница 44: ...P N XXXXXXXX SERVICE MANUAL Valor 1000W SERIES SCALES...