1-1
CHAPTER 1 INTRODUCTION
1. INTRODUCTION
This service manual contains instructions for the repair and maintenance work to be performed by Ohaus
Dealers or Ohaus authorized service centers. Knowledge of the operation of the Scale is assumed.
Instruction manuals are included with this service manual. For complete information on operation, refer
to the Instruction Manual.
This manual covers maintenance on the following:
Trooper
®
Scales (capacities 3kg, 6kg, 15kg and 30kg.)
Trooper
®
Count Scales (capacities 3kg, 6kg, 15kg and 30kg.)
The contents of this manual are contained in five chapters and an Appendix with calibration information.
Chapter 1 Introduction
- Contains information about service facilities, tools and test equipment, test
masses, and service strategy.
Chapter 2 Diagnosis
- Contains information on problem verification, scale examination, preliminary
checks, troubleshooting tables, interconnection diagrams and wiring diagrams.
Chapter 3 Repair Procedures
- Contains detailed repair procedures for all major components.
Chapter 4 Testing
- Contains performance tests, specifications and communications testing procedures.
Chapter 5 Parts Lists
- Contains exploded views of Trooper Scale and Trooper Count Scale identifying all
serviceable replacement components with associated parts lists.
Appendix A Calibration
- Contains linearity calibration procedures for the Trooper Scale and Trooper
Count Scale.
Appendix B Trooper Count Scale Tools
- Contains information on scale configuration, library data and
testing output commands from a computer.
1.1 SERVICE FACILITIES
The service area should be a stable environment.
The bench area should be clean and should contain an antistatic mat with a personnel grounding clip to
protect internal circuit boards. The ideal electrical power source for the scales should be a dedicated line
to avoid sudden fluctuations or voltage drops caused by external equipment drawing heavy current.
The service area for the scales should be away from direct sunlight, overhead heating or air conditioning
ducts, magnetic fields such as motors or large transformers or near vibrating sources such as machinery,
motors or vibrating equipment.
The power outlet should be grounded for safety. Sufficient space should be provided around the scale as
not to be affected by other equipment. This will ensure that the scale is operated under ideal conditions.