B-7
APPENDIX B - TROOPER COUNT SCALE TOOLS
B.3.3 Test Command
The Test Command Screen enables the user to send (type in) a test command from the computer keyboard
and log the data output on-screen. Most commands mimic the button functions of the scale.
Initiate use of the Test Command screen by selecting the [Connect] button first to establish connection
between the scale and the computer.
Note
: A communication error message will appear if connection cannot be established (refer to Sections
B.4: Troubleshooting and B.3.4: COM Port Screen for further details).
Type in a test command followed by the [Enter] key
(test commands are case-sensitive, use upper-case
letters only)
. The available Test Commands are listed and described as follows (for examples, refer to Test
Command Screen):
• P (Print)
P (Print)
P (Print)
P (Print)
P (Print)
Displays the print data output of the scale on the “Response from scale” field.
• Z (Zero)
Z (Zero)
Z (Zero)
Z (Zero)
Z (Zero)
Zeroes the scale (similar to pressing the [ON/ZERO
Off] button from the scale). A result of “OK”
on-screen confirms the operation. The scale will also reflect that the operation was executed.
• T (Tare)
T (Tare)
T (Tare)
T (Tare)
T (Tare)
Tares the scale (similar to pressing the [TARE] button from the scale). A result of “OK” on-screen
confirms the operation. The scale will also reflect that the operation was executed.
• V (Get Version Number)
V (Get Version Number)
V (Get Version Number)
V (Get Version Number)
V (Get Version Number)
Displays the model and software version of the Trooper Count Scale on the “Response from
scale” field.