1-3
CHAPTER 1 INTRODUCTION
1.4 TOOLS AND TEST EQUIPMENT REQUIRED
In order to properly service the Ohaus Trooper Scales, a standard electronic tool kit is required. No
special tools and test items are required.
1.4.1 Special Tools
None required.
1.4.2 Standard Tools and Test Equipment
1. Digital Voltmeter (DVM) - Input impedance of at least 10 megohms in the 1 Volt dc position.
2. Standard Electronics tool kit
1.5 TEST MASSES REQUIRED
The masses required to test the Ohaus Trooper Scales must meet the requirements of ASTM
Class 4 Tolerance. The mass values are listed in Table 1-1.
TABLE 1-1. TROOPER CALIBRATION MASSES
Cal in kg:
Span cal choices
Linearity cal (fixed)
TR3RS
1, 2, 3kg
2 & 3 kg
TR6RS
2, 4, 6kg
4 & 6 kg
TR15RS
5, 10, 15kg
10 & 15 kg
TR30RS
10, 20, 30kg
20 & 30 kg
Cal in lb:
Span cal choices
Linearity cal (fixed)
TR3RS
2, 4, 6 lb
4 & 6 lb
TR6RS
5, 10, 15 lb
10 & 15 lb
TR15RS
10, 20, 30 lb
20 & 30 lb
TR30RS
20, 40, 60 lb
40 & 60 lb