4-7
CHAPTER 4 MAINTENANCE PROCEDURES
4.2.5 Calibration (Cont.)
Procedure
TURNING CALIBRATION LOCK SWITCH ON
To turn the software lock on:
Press and hold the G/N/T/
MENU button until
MENU appears. Release it and SETuP appears.
(if already in the main menu, skip to the next
step).
Repeatedly press the PRINT/
UNITS button until
LOCSW is displayed.
Repeatedly press the G/N/T/
MENU button until
LCLOFF is displayed.
Press the PRINT/
UNITS button to scroll through
the choices until LCLON appears.
Press the G/N/T/
MENU button repeatedly until
the scale returns to weighing
The scale is now ready for weighing.