CHAPTER 1
GETTING STARTED
Ohaus Corporation www.ohaus.com
1-18
7000 Series T71 Indicator Service Manual
COmM 1
BAUD
COmM 1
PARITY
COmM 1
STOP BIT
COmM 1
HANDSHAKE
COmM 1
ALT COMMAND
1.9 RS232 INTERFACE CONNECTION
The T71 has a bi-directional RS232 interface for communication with printers and computers.
When the Indicator is connected to a printer or Programmable Logic Controller (PLC), displayed
data can be output various ways, depending on PRINT menu settings.
For hardware setup of the RS232 Interface, see Section 1.5.2, page 1-6.
1.9.1 Setting Communications Parameters (COM1)
Note:
Only the COM1 menu appears unless the COM2 interface is set up.
The T71 Indicator’s default communication parameters are baud rate 9600 bits per second
(bps), 8 databits, no parity
,
Handshake XON/XOFF, Address Off, Alt Command (for Print) Off,
Alt Tare Off, and Alt Zero Off.
To change these settings use the COM1 menu, as follows:
1. Press and hold
TARE/Menu
until LFT OFF flashes and CALIBRATION
appears. Press
No
until
COM1
appears. Press
Yes.
RESEt
appears. Press
No
(unless you want to restore the factory defaults).
2. COM
1
bAUd
appears. Press
Yes
.
3. The default Baud rate setting,
9600
bps, appears. To change the setting, press
No.
The menu cycles to the next setting each time you press No: first 9600, then 300, 600,
1200, 2400, 4800, and 9600 again. Press
Yes
when the desired setting appears.
COM 1
bAUd
appears again. Press
No
.
4. COM
1
PAritY
appears. Press
Yes
. The default Parity setting,
8 NONE
, appears. To change the setting, press
No
.
The menu cycles to the next setting each time you press
No
:
first 7 Even, then 7 Odd, 7 None, and 8 None again. Press
Yes
to accept a setting.
PAritY
appears again. Press
No
.
5. COM
1
StOP BIT
appears. Press
Yes
. The default Stop Bit
setting,
1
, appears. To change the setting, press
No
.
The menu toggles to the next setting each time you press No:
first 1, then 2, then 1 again. Press
Yes
when the desired setting
appears. COM 1
StOP BIT
appears again. Press
No
.
6. COM
1
HANdSHAKE
appears. Press
Yes
. The default setting,
NONE
, appears. To change the setting, press
No
.
The menu toggles to the next setting each time you press No:
first
XON-XOFF
, then
HARDWARE,
then
NONE
again. Press
Yes
to accept a setting. Then menu moves to the next item.
7.
COM 1
ALT COMMAND
appears. Press
Yes
to set a different
command character for the P (Print), T (Tare) and Z (Zero)
commands. Otherwise, press
No
. Then press
Yes
when
COM 1 END COM1
appears.
At this point, if a second communication interface is present, the COM2 menu is available.
Otherwise, press Exit to return to current application mode.
Содержание T71P
Страница 2: ......
Страница 4: ......
Страница 28: ...CHAPTER 1 GETTING STARTED Ohaus Corporation www ohaus com 1 22 7000 Series T71 Indicator Service Manual ...
Страница 44: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 6 7000 Series T71 Indicators Service Manual ...
Страница 47: ......
Страница 48: ... 80252588 P N 80252588 SERVICE MANUAL 7000 SERIES T71 INDICATORS ...