CHAPTER 1
GETTING STARTED
Ohaus Corporation www.ohaus.com
1-14
7000 Series T71 Indicator Service Manual
SPAN
kg
10
SPAN
lb
5
SPAN
kg
5
SPAN
lb
--C--
SPAN
lb
0
0.000
lb
SPAN
lb
DONE
SPAN
lb
--C--
1.8.2 Span
Calibration
Use this calibration method to adjust the zero calibration point and span calibration point,
without affecting the linearity calibration. Be sure there is no load on the scale.
Press and hold
TARE/Menu
until LFT OFF flashes and CALIBRATION
appears. Press
Yes
. CAL
ZERO
appears. Press
No
.
CAL
SPAN
appears. Press
Yes
. SPAN and the current unit of measure
and Span calibration point flash.
Note
: To change the span calibration point, press
No
and enter the
value using the keypad. Then press
Yes
.
To change the calibration unit of measure, press
No
to alternate
between kg and lb.
Place the specified calibration weight on the scale and press
Yes
.
The display shows SPAN
--C--
, followed by SPAN
DONE
.
SPAN
0
appears. With no load on the scale, press
Yes
.
The display shows SPAN
--C--
, followed by SPAN
DONE,
and returns to
the current application mode.
Содержание T71P
Страница 2: ......
Страница 4: ......
Страница 28: ...CHAPTER 1 GETTING STARTED Ohaus Corporation www ohaus com 1 22 7000 Series T71 Indicator Service Manual ...
Страница 44: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 6 7000 Series T71 Indicators Service Manual ...
Страница 47: ......
Страница 48: ... 80252588 P N 80252588 SERVICE MANUAL 7000 SERIES T71 INDICATORS ...