CHAPTER 1
GETTING STARTED
Ohaus Corporation www.ohaus.com
1-6
7000 Series T71 Indicator Service Manual
1.6.2 Jumper
Connections
When a 4-wire load cell is attached to terminal block J4, jumpers W1 and W2 must be shorted.
When a 6-wire load cell is attached to terminal block J4, jumpers W1 and W2 must be open.
When a 4-wire load cell is attached to terminal block J7, jumpers W3 and W4 must be shorted.
When a 6-wire load cell is attached to terminal block J7, jumpers W3 and W4 must be removed.
+EXC +SIG -SIG -EXC
+SENS GND -SENS
Figure 1-5. Load Cell wiring.
Pin Connection
1 +Excitation
2 +Sense
3 +Signal
4 GND
5 –Signal
6 –Sense
7 –Excitation
Figure 1-6. Open/shorted jumpers.
1.6.3 RS232 Interface
The T71P has a built-in RS232 Interface: a standard 9-pin D-style female connector on the back
of the T71P Housing. In addition, each model (T71P and T71XW), has an optional COM2
connector, which can be either RS232 or RS422/485. The optional interface is available
separately, with its own instructional manual for installation and configuration.
The RS232 pin configuration for the T71P is illustrated in Figure 1-7, below.
1
2
3
4
5
6
7
8
9
Figure 1-7. RS232 Pins.
Pin Connection
1 N/C
2 TXD
3 RXD
4 N/C
5 GND
6 N/C
7 CTS
8 RTS
9 N/C
For software setup of the RS232 Interface, through the Indicator’s menu, see Section 1.9.
All other aspects of hardware configuration and setup are described in the Instruction Manual.
Содержание T71P
Страница 2: ......
Страница 4: ......
Страница 28: ...CHAPTER 1 GETTING STARTED Ohaus Corporation www ohaus com 1 22 7000 Series T71 Indicator Service Manual ...
Страница 44: ...CHAPTER 5 PARTS LISTS DIAGRAMS Ohaus Corporation www ohaus com 5 6 7000 Series T71 Indicators Service Manual ...
Страница 47: ......
Страница 48: ... 80252588 P N 80252588 SERVICE MANUAL 7000 SERIES T71 INDICATORS ...