STARTER 2100 Bench pH Meters EN-11
Press button-
to reject the calibration and return to the measurement screen.
Press button-
to do the 2-point calibration.
Note:
With the 1-point calibration only the
offset
is adjusted. If the pH electrode was previously
calibrated with 2-point calibration the previously stored
slope
will remain. Otherwise theoretical
100% slope
(-59.16 mV / pH @25 °C) will be used.
4.1.3 Performing 2-point calibration
Perform 1-point calibration as described above.
Rinse the pH electrode with pure water and wipe off water with a tissue.
Place the electrode in the next calibration buffer stir and wait, then press button-
,
“
Cal
2
” displays on the top right of the screen and is blinking. appears and is blinking
during calibration.
When reading is stable (normally the reading does not change in 5s), press button-
to lock the reading and finish 2-point calibration, disappears. The buffer pH value (e.g.
4.01 pH
) with the temperature display on the screen.
The 2-point calibration is finished, now we have 2 options:
Press button-
to store the 2-point calibration, the
offset
and
slope
are shown on the
display for 3 seconds then returns to the measurement screen.
Press button-
to reject the calibration and return to the measurement screen.
Note:
The use of a temperature electrode (e.g. STTEMP30) or a pH electrode with a built-in
temperature sensor (3-in-1 pH electrode) is recommended. If you use the
MTC
mode, you should
enter the correct temperature value and keep all buffer and sample solutions at the set
temperature. (See 3.1)