STARTER 3100C Bench Conductivity Meter EN-13
The details for the second line are:
Auto EP, 85.1
μ
S/cm, 26.8°C, ATC
| | | |---- Auto Temp. Compensation
| | |----------------Temperature value and unit
| |------------------------- Conductivity value
|---------------------------------------- Auto End Point
The details for the third line are:
Ref.25°C , T.Comp.Coeff. 2.00%/°C
| |
| |---------------- Temperature compensation coefficient value
|------------------------------------------- Reference temperature, 20 or 25 °C
The format for the print-out following a TDS measurement is:
End Point, Value, Temp., ATC/MTC
Manual EP, 38.0mg/L, 23.2°C , ATC
Ref.25°C , T.Comp.Coeff. 2.00%/°C
TDS-F 0.5
The format for the print-out following a Salinity measurement is:
End Point, Value, Temp., ATC/MTC
Manual EP, 0.08psu, 25.0°C , MTC
The print-out in case of an error message is:
End Point, Value, Temp., ATC/MTC
Error4
The print-out for a successful calibration is:
Sta
ndard: 1413
μ
S/cm
Temperature: 30.0 °C
Calibration value: 1552
μ
S/cm
Cell Constant: 1.66 cm-1
ATC/MTC: ATC
Printing from memory When scrolling through the memory you can print the entry that is currently
viewed by pressing and holding
for 2 seconds. The printout format is followings:
Recall 08:
EndPoint, Value, Temp., ATC/MTC
Auto EP, 68.8
μ
S/cm, 26.8°C , ATC
Ref.25°C , T.Comp.Coeff. 2.00%/°C
www.
GlobalTestSupply
.com
Find Quality Products Online at: