EN-8 STARTER 3100C Bench Conductivity Meter
3 SETUP
3.1 Set
temperature
unit
x
Power the meter on by pressing
.
x
Press and hold
until the set up icon
appears on the display and the current
temperature unit blinks. (°C or °F)
x
Use
or
buttons to select temperature unit .
x
Press
to confirm your setting. Continue with temperature value setting by
following or press
to go back to the measurement screen.
Note: °C = 5/9 (°F -32)
3.2 Set
MTC
temperature
value
After the temperature unit setting, the meter goes to setup manual temperature (MTC) value
setting.
Use
or
buttons to adjust the temperature value.
Press
to confirm your setting. Continue with calibration standard setting by following or
press
to go back to the measurement screen.
3.3 Set
calibration
standard
After the MTC temperature value setting, the meter goes to setup calibration standard.
Use
or
buttons to select calibration standard.
Press
to confirm your setting. Continue with temperature correction coefficient setting
by following or press
to go back to the measurement screen.
3.4
Set temperature correction coefficient
After the calibration standard setting, the meter goes to the temperature correction coefficient