EN-12 STARTER 3100C Bench Conductivity Meter
4.4
Using the memory
4.4.1 Storing
a
reading
The
STARTER 3100C
can store up to 99 endpoint results.
Press
when the measurement reaches endpoint.
M01
indicates that one result has
been stored.
If you press
when
M99
is displayed,
FUL
displays to indicate the memory is full. To store
further data you will have to clear the memory. (See 3.4.3)
4.4.2 Recalling
from
memory
Press and hold
to recall the stored values from memory when the current
measurement reaches endpoint.
Press
or
to scroll through the stored results.
R01
to
R99
indicates which result
is being displayed.
Press
to exit.
4.4.3
Clearing the memory
Pressing
or
to scroll through the stored results until
“
MRCL
”
appears.
Press
,
CLr
blinks;
Now we can two choices:
Press
to confirm the deletion of all the stored data.
Press
to return to the measurement mode without deleting the memory.
4.5 Printing
If a printer (e.g. SF-F40A) is connected to the STARTER 3100C, a print-out is automatically
generated after each end pointed measurement or calibration. (printer baud rate 9600bps; 8 data
bit; none parity bit; 1 stop bit)
The format for the print-out following a Cond. measurement is:
End Point, Value, Temp., ATC/MTC
Auto EP, 68.8
μ
S/cm, 26.8°C , ATC
Ref.25°C , T.Comp.Coeff. 2.00%/°C