CHAPTER 1 GETTING STARTED
Starter Series Portable Meter Service Manual
1-5
Ohaus Corporation www.ohaus.com
1.7 METER OPERATION
– STARTER ST300
This section contains information on the basic operation of the Starter ST300 meter.
1.7.1 OVERVIEW OF THE CONTROLS
Displays
Figure 1-1. ST300 display
1
Electrode condition
Slope: more than 95%
and offset: ± (0-15) mV
Electrode condition is
good
Slope: 90-95%
or offset: ± (15-35)
mV
Electrode condition is
acceptable
Slope: less than 90%
or offset: ± (35-60) mV
Electrode condition is not good
or needs cleaning
2 Endpoint stability icon
; Auto endpoint icon
3
Measurement icon - ; measurement or calibration is running
4
Calibration icon -
Ca
l ; calibration in progress
5
Setup icon -
; instrument is in the setup mode
6
Battery status icon shows the condition of the batteries - fully charged, half-charged or
fully discharge.
7
Temperature during measurement or
Offset
(mV) in calibration
8
Calibration point
Cal
/ Buffer group
/Memory number
MR
/ Error index
Err
9
Auto temperature compensation -
ATC
; Manual temperature compensation -
MTC
10 Temperature during measurement or offset in calibration process