CHAPTER 4 PARTS LISTS & DIAGRAMS
Starter Series Portable Series Service Manual 4-1
Ohaus Corporation www.ohaus.com
This section of the manual contains exploded views of the Starter ST300, ST300C, ST300D
and ST400D portable meter. The exploded view drawings are designed to identify the parts
which can be serviced on the portable meter in the field.
NOTE:
Attention
: In all cases where a part is replaced, the meter must be thoroughly checked after
the replacement is made. The meter
MUST
meet the parameters of all applicable
specifications in this manual.
Step to confirm the meter operates within specification:
1. Obtain a known good working condition electrode and buffer solution in hand.
2. Link the electrode to the portable meter, perform a calibration. After completed the
calibration do a measurement and compare with the meter specification.
If further technical information is needed, please contact your local Ohaus distributor, or:
www.ohaus.com
Ohaus Corporation
7 Campus Drive
Suite 310
Parsippany, NJ 07054 USA
Tel: +1 973-377-9000
Fax: +1 973-593-0359
In the United States call toll free, 800-526-0659 between 8:00 a.m. and 6:00 p.m. EST.