background image

EN-16

 SCOUT

TM

 SERIES BALANCES - SPX

6. TECHNICAL DATA

The technical data is valid under the following ambient conditions: 
Indoor use only 
Operating temperature range: +5°C to +40°C 
Relative humidity: 10% to 80% at 31°C, decreasing linearly to 50% at 40°C, non-condensing 
Altitude: Up to 2000 m 
Power: AC power adaptor input 100-240V 50/60 Hz and output 5 V DC 1 A, or 4 AA batteries 
Pollution degree: 2 
Installation category: II 
Main supply voltage fluctuations: up to ± 10% of the nominal voltage 

6.1 Specifications

 

TABLE 6-1. SPECIFICATIONS 

Model 

SPX123 

SPX223 

SPX222 

SPX422 

SPX622 

SPX1202 

SPX2202 

Capacity (g) 

120 

220 

220 

420 

620 

1200 

2200 

Readability (g) 

0.001 

0.001 

0.01 

0.01 

0.01 

0.01 

0.01 

Repeatability (Std. Dev.) (g) 

0.002 

0.002 

0.01 

0.01 

0.01 

0.02 

0.02 

Linearity (g) 

0.003 

0.003 

0.01 

0.01 

0.02 

0.03 

0.03 

Span Calibration Mass* 

100 g 

200 g 

200 g 

200 g 

300 g 

1 kg 

2 kg 

Linearity Calibration Mass 

50, 100 g 

100, 200 g  100, 200 g  200, 400 g  300, 600 g 

500 g, 

1 kg 

1 kg, 2 kg 

Stabilization Time (s) 

1.5 

1.5 

Construction 

ABS plastic housing with 304 stainless steel (SST) pan 

Draftshield 

Yes 

No 

Calibration 

User-selectable external span or linearity calibration / Digital with external weight 

Tare Range 

Full Capacity by subtraction 

Weighing Units** 

g, kg, ct, N, oz, ozt, dwt, lb, lb:oz, grn,Tael (HongKong),Tael (Singapore),Tael (Taiwan), tical, 

tola, C (Custom unit) 

Application Modes 

Weighing, Parts Counting, Percent Weighing, Check Weighing, Totalization, Display Hold 

Typical Battery Life 

80 hours 

120 hours 

80 hours 

Specified Temperature 
Range 

10° C (50° F) to 40° C (104° F) at 10% to 80% relative humidity, non-condensing 

Storage Conditions 

-20° C (-4° F) to 55° C (131° F) at 10% to 90% relative humidity, non-condensing

Communication 

RS232, USB Host, USB Device, Ethernet or Bluetooth*** (all available as accessory) 

Display Type 

Backlit LCD: 6-digit 7-segment with white LED backlight 

Display Size 

0.78 in / 20 mm digits 

Pan Size (W x D) 

Ø 93 mm / 3.7 in

Ø 120 mm / 4.7 in

170 x 140 mm / 

6.7 x 5.5 in 

Balance Dimensions 
(W x D x H) 

202 x 222 x 103 mm / 

8 x 8.7 x 4.1 in 

202 x 224 x 54 mm / 8 x 8.8 x 2.1 in 

Shipping Dimensions 
(W x D x H) 

300 x 250 x 129mm / 

11.8 x 9.8 x 5.1 in 

300 x 250 x 86 mm / 11.8 x 9.8 x 3.4 in 

Net Weight 

1 kg / 2.2 lb 

Shipping Weight 

1.5 kg / 3.3 lb 

Notes: 

* Calibration weights are included with models up to 620g capacity.
** Availability is dependent on model and region.
*** Bluetooth kit is only available in certain regions according to the local regulations

(866) 433-6682 • (281) 359-8538 • [email protected] • www.novatech-usa.com

Содержание Scout SPX1202

Страница 1: ...Scout TM Series Balances SPX Instruction Manual sales novatech usa com www novatech usa com Tel 866 433 6682 Fax 866 433 6684 Tel 281 359 8538 Fax 281 359 0084...

Страница 2: ...comply with these warnings could result in personal injury and or property damage Retain all instructions for future reference Verify that the local AC power supply is within the input voltage range p...

Страница 3: ...ocation For best performance the Scout SPX balance should be used in a clean stable environment Do not use the balance in environments with excessive drafts with rapid temperature changes near magneti...

Страница 4: ...mpartment Note After power on it is recommended to warm up the balance for at least 5 minutes before using it 2 5 Initial Calibration When the Balance is first installed and when it is moved to anothe...

Страница 5: ...o Off Turns the balance Off Units Changes the weighing unit Mode Allows changing the application mode Menu Enter the User menu Menu Function Short Press Yes Accepts the current setting on the display...

Страница 6: ...nd hold the On Zero Off button until OFF is displayed 3 3 Calibration Operation When the balance is operated for the first time a span calibration is recommended to ensure accurate weighing results Be...

Страница 7: ...al pieces until the desired count is reached 8 To clear the stored APW press and hold Mode until Count is displayed Press Yes when Clr pwW is displayed Note Press Function to view the current APW 3 6...

Страница 8: ...der icon will light If the sample is within the target weight range the accept symbol will light If the sample is over the target weight range over icon will light Note Press Function to view the low...

Страница 9: ...clear the stored total press and hold Mode until total is displayed When Clr tot is displayed press Yes 3 9 Hold Mode There are two modes for the display hold Peak Hold allows the user to capture and...

Страница 10: ...Peak Off Reset g kg ct N oz ozt dwt lb lb oz grn hkt sgt twt t tola tical C End End End End Notes Only available in certain regions When the Security Switch is in locked position the Calibration funct...

Страница 11: ...hange of 3 divisions per second has been exceeded Stable Range Stable Set the amount the reading can vary while the stability symbol remains on 0 5d 0 5 balance division 1d 1 balance division 2d 2 bal...

Страница 12: ...available for use with the Mode button Reset reset Reset the Mode menu to factory defaults NO not reset YES reset Weigh wWEIGH Set the status OFF disabled ON enabled Count COUNt Set the status OFF di...

Страница 13: ...ts in the menu must be turned on to be active Note Available units vary by model and local regulations Custom Unit C Use the Custom Unit to display weight in an alternative unit of measure The custom...

Страница 14: ...2 rESEt will be displayed Press No Unit g will be displayed 3 Press No to toggle through active units until Unit c is displayed 4 Press Yes On c will be displayed then press Yes 5 Factor is displayed...

Страница 15: ...is level and secure and that the transportation lock has been released Power on the balance then use a string or wire to attach items to be weighed Figure 4 3 Weigh below application Connecting the I...

Страница 16: ...ght 1d Balance shows error then exits parts counting or goes to Clr APU Lo ref Percent Reference weight or APW is too low for accurate results CAL E Fail to do calibration USB Err Cannot find menu or...

Страница 17: ...ration Digital with external weight Tare Range Full Capacity by subtraction Weighing Units g kg ct N oz ozt dwt lb lb oz grn Tael HongKong Tael Singapore Tael Taiwan tical tola C Custom unit Applicati...

Страница 18: ...200 g 300 g 2 kg 5 kg 8 kg Linearity Calibration Mass 200 400 g 300 600 g 1 kg 2 kg 3 kg 6 kg 4 kg 8 kg Stabilization Time s 1 Construction ABS plastic housing with 304 stainless steel SST pan Drafts...

Страница 19: ...0 0005 4 8500 x 0 0005 13 6685 x 0 0005 18 080 x 0 005 Pound Ounce lb oz 1lb 5 870oz x 0 005oz 4lb 13 600oz x 0 005oz 13lb 10 700oz x 0 005oz 18lb 1 25oz x 0 05oz Grain grn 6480 x 2 9570 x 2 33950 x...

Страница 20: ...or electrical and electronic equipment If you have any questions please contact the responsible authority or the distributor from which you purchased this device Should this device be passed on to oth...

Страница 21: ...es to be defective at no charge provided that the product is returned freight prepaid to OHAUS This warranty does not apply if the product has been damaged by accident or misuse exposed to radioactive...

Страница 22: ...ltweite Gesh ftsstellen Con uffici in tutto il mondo www ohaus com 30268991 P N 30268991 C 2016 OHAUS Corporation all rights reserved todos los derechos reservados tous droits r serv s Alle Rechte vor...

Отзывы: