background image

 

SCOUT

TM

 SERIES BALANCES - SPX                                                                                              EN-3 

2.4 Connecting Power

 

AC Adapter Installation 

AC power is used to power the scale when battery power is not needed. First, connect the AC 
Adapter (supplied) to the AC Adapter Input Jack at the rear of the balance then connect the AC 
plug to an electrical outlet. 

 
                                                                               
 

 

 
 
 

 

 

Figure 2-4. Rear and bottom view of balance  

 

Battery Installation 

Install the four “AA” batteries with polarity as shown in the battery compartment. 
 

Note:  

After power on, it is recommended to warm up the balance for at least 5 minutes before using it. 
 

 

2.5 Initial Calibration

 

When the Balance is first installed, and when it is moved to another location, it must be 
calibrated to ensure accurate weighing results. Have the appropriate calibration masses 
available before beginning calibration. Refer to the Calibration Section for masses and 
calibration procedure.  
 

 
  
 
 

 
 
 
 
 
 
 

AC Adapter Input Jack 

Battery Compartment Cover

 

Interface Connection 

Security Switch

 

Weigh Below Hole

 

Security slot (Anti-theft lock slot) 

(866) 433-6682 • (281) 359-8538 • [email protected] • www.novatech-usa.com

Содержание Scout SPX1202

Страница 1: ...Scout TM Series Balances SPX Instruction Manual sales novatech usa com www novatech usa com Tel 866 433 6682 Fax 866 433 6684 Tel 281 359 8538 Fax 281 359 0084...

Страница 2: ...comply with these warnings could result in personal injury and or property damage Retain all instructions for future reference Verify that the local AC power supply is within the input voltage range p...

Страница 3: ...ocation For best performance the Scout SPX balance should be used in a clean stable environment Do not use the balance in environments with excessive drafts with rapid temperature changes near magneti...

Страница 4: ...mpartment Note After power on it is recommended to warm up the balance for at least 5 minutes before using it 2 5 Initial Calibration When the Balance is first installed and when it is moved to anothe...

Страница 5: ...o Off Turns the balance Off Units Changes the weighing unit Mode Allows changing the application mode Menu Enter the User menu Menu Function Short Press Yes Accepts the current setting on the display...

Страница 6: ...nd hold the On Zero Off button until OFF is displayed 3 3 Calibration Operation When the balance is operated for the first time a span calibration is recommended to ensure accurate weighing results Be...

Страница 7: ...al pieces until the desired count is reached 8 To clear the stored APW press and hold Mode until Count is displayed Press Yes when Clr pwW is displayed Note Press Function to view the current APW 3 6...

Страница 8: ...der icon will light If the sample is within the target weight range the accept symbol will light If the sample is over the target weight range over icon will light Note Press Function to view the low...

Страница 9: ...clear the stored total press and hold Mode until total is displayed When Clr tot is displayed press Yes 3 9 Hold Mode There are two modes for the display hold Peak Hold allows the user to capture and...

Страница 10: ...Peak Off Reset g kg ct N oz ozt dwt lb lb oz grn hkt sgt twt t tola tical C End End End End Notes Only available in certain regions When the Security Switch is in locked position the Calibration funct...

Страница 11: ...hange of 3 divisions per second has been exceeded Stable Range Stable Set the amount the reading can vary while the stability symbol remains on 0 5d 0 5 balance division 1d 1 balance division 2d 2 bal...

Страница 12: ...available for use with the Mode button Reset reset Reset the Mode menu to factory defaults NO not reset YES reset Weigh wWEIGH Set the status OFF disabled ON enabled Count COUNt Set the status OFF di...

Страница 13: ...ts in the menu must be turned on to be active Note Available units vary by model and local regulations Custom Unit C Use the Custom Unit to display weight in an alternative unit of measure The custom...

Страница 14: ...2 rESEt will be displayed Press No Unit g will be displayed 3 Press No to toggle through active units until Unit c is displayed 4 Press Yes On c will be displayed then press Yes 5 Factor is displayed...

Страница 15: ...is level and secure and that the transportation lock has been released Power on the balance then use a string or wire to attach items to be weighed Figure 4 3 Weigh below application Connecting the I...

Страница 16: ...ght 1d Balance shows error then exits parts counting or goes to Clr APU Lo ref Percent Reference weight or APW is too low for accurate results CAL E Fail to do calibration USB Err Cannot find menu or...

Страница 17: ...ration Digital with external weight Tare Range Full Capacity by subtraction Weighing Units g kg ct N oz ozt dwt lb lb oz grn Tael HongKong Tael Singapore Tael Taiwan tical tola C Custom unit Applicati...

Страница 18: ...200 g 300 g 2 kg 5 kg 8 kg Linearity Calibration Mass 200 400 g 300 600 g 1 kg 2 kg 3 kg 6 kg 4 kg 8 kg Stabilization Time s 1 Construction ABS plastic housing with 304 stainless steel SST pan Drafts...

Страница 19: ...0 0005 4 8500 x 0 0005 13 6685 x 0 0005 18 080 x 0 005 Pound Ounce lb oz 1lb 5 870oz x 0 005oz 4lb 13 600oz x 0 005oz 13lb 10 700oz x 0 005oz 18lb 1 25oz x 0 05oz Grain grn 6480 x 2 9570 x 2 33950 x...

Страница 20: ...or electrical and electronic equipment If you have any questions please contact the responsible authority or the distributor from which you purchased this device Should this device be passed on to oth...

Страница 21: ...es to be defective at no charge provided that the product is returned freight prepaid to OHAUS This warranty does not apply if the product has been damaged by accident or misuse exposed to radioactive...

Страница 22: ...ltweite Gesh ftsstellen Con uffici in tutto il mondo www ohaus com 30268991 P N 30268991 C 2016 OHAUS Corporation all rights reserved todos los derechos reservados tous droits r serv s Alle Rechte vor...

Отзывы: