EN-6
Ethernet Interface
Input:
The following interface commands will be acknowledged by the Balance.
They are case sensitive.
The balance will return “ES” for invalid commands.
Command
Function
IP
Immediate Print of displayed weight (stable or unstable).
P
Print displayed weight (stable or unstable).
CP
Continuous Print.
SP
Print on Stability.
SLP
Auto Print stable non-zero displayed weight.
SLZP
Auto Print stable non-zero weight and stable zero reading.
xP
Interval Print x = Print Interval (1-3600 sec) 0P ends interval Print
0P
Turn off Auto Print
H
H x “text”
Enter Header line , where x = line number 1 to 5, “text” = header
text up to 24 alphanumeric characters
F
F x “text”
Enter Footer line , where x = line number 1 to 2, “text” = footer
text up to 24 alphanumeric characters
Z
Same as pressing Zero Key
T
Same as pressing Tare Key.
xT
Establish a preset Tare value in displayed unit. x = preset tare value.
Sending 0T clears tare (if allowed).
PT
Prints Tare weight stored in memory.
PM
Print current application mode (weighing mode).
xM
Set current application mode to x. x depends on application
1M: WEIGH, 2M: COUNT,3M: PERCENT, 4M: CHECK, 5M: DYNAMIC, 6M:
TOTAL, 7M:DENSITY, 8M: HOLD, 9M:MOLE
M
Scroll to the next enabled mode.
PU
Print Current weighing unit: g, kg, lb, oz, e
tc….
xU
Set balance to unit x: g, kg etc.
1U: g, 2U: kg,3U: ct, 4U: N, 5U: oz, 6U: ozt, 7U: dwt, 8U: lb, 9U:lb:oz, 10U:
grn, 11U: thk, 12U: tsg, 13U:ttw , 14U: tola, 15U:c
U
Scroll to the next enabled unit.
ON
Brings out of Standby
OFF
Goes to Standby.
C
Begin Span Calibration
AC
Abort Calibration.
PSN
Print Serial Number.
PV
Print Version: print name, software revision and LFT ON (if LFT is set ON).
x#
Set Counting APW (x) in grams. (must have APW stored)
P#
Print Counting application APW.
x%
Set Percent application reference weight (x) in grams. (must have reference
weight stored)
P%
Print Percent application reference weight.
xCO
Set Checkweighing Over Limit in grams x.
xCU
Set Checkweighing Under Limit in grams x.
PCO
Print Checkweighing Over Limit.
PCU
Print Checkweighing Under Limit.
xMM
Set Molar Mass in g/mol .
PMM
Print Molar Mass
xS
0 = print unstable data, 1 = print stable only
xFMT
0
= New Scout print format (default) ; 1 = Scout Pro print format 1;
2 = Scout Pro print format 2; 3 = for certain POS system.
xRL
0 = disable response; 1 = enable response.
www.
GlobalTestSupply
.com
Find Quality Products Online at:
Содержание Scout 30268986
Страница 11: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 51: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 61: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 81: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...