EN-14
RE Series
6.
Span Calibration Sequence:
Display Description
Actions
Lb
Calibration Unit:
kg or lb
Press [FIX]
to select desired unit and proceed to the
next step.
If needing to switch between kg and lb,
press
[>T<]
.
CAP 30
Scale Capacity
(in the Calibration Unit
selected)
Press [FIX]
to select the displayed setting and
proceed to the next step. Make sure the displayed
value matches the scale model’s capacity.
Do not press
[>T<]
, which will toggle through other
capacity settings that may not match the scale
model’s.
------
Establish zero
Ensure the pan is empty and the scale stable.
Press [FIX]
to initiate countdown sequence.
5 ... 0
Countdown
The countdown will repeat if the scale is not stable
during this process.
Add20
2/3 calibration point
Press [>T<]
to switch between 2/3 or full capacity,
then
Place the indicated weight on the pan, then
press [FIX]
to proceed to the next step.
5 ... 0
Countdown
The countdown will repeat if the scale is not stable
during this process.
Save
Save or Abort
Remove the weight from the pan, then press [FIX]
to save settings and restart scale to weighing mode.
If needing to abort calibration and retain previous
calibration data, press
[>T<]
(“Abort” will be
displayed) then
[FIX]
at this time (or turn the scale
off at any time during the process).