EN-14 

                      RE Series 

 

 

6.

 

Span Calibration Sequence: 

 

Display Description 

Actions 

Lb 

Calibration Unit:  
kg or lb 

Press [FIX]

 to select desired unit and proceed to the 

next step. 
If needing to switch between kg and lb, 

press 

[>T<]

CAP 30 

Scale Capacity  
(in the Calibration Unit 
selected) 

Press [FIX]

 to select the displayed setting and 

proceed to the next step.  Make sure the displayed 
value matches the scale model’s capacity. 
Do not press 

[>T<]

, which will toggle through other 

capacity settings that may not match the scale 
model’s. 

------ 

Establish zero 

Ensure the pan is empty and the scale stable. 

Press [FIX]

 to initiate countdown sequence. 

5 ... 0 

Countdown 

The countdown will repeat if the scale is not stable 
during this process. 

Add20 

2/3 calibration point 

Press [>T<]

 to switch between 2/3 or full capacity, 

then 

Place the indicated weight on the pan, then 

press [FIX]

 to proceed to the next step. 

5 ... 0 

Countdown 

The countdown will repeat if the scale is not stable 
during this process. 

Save 

Save or Abort 

Remove the weight from the pan, then press [FIX]

 

to save settings and restart scale to weighing mode. 
If needing to abort calibration and retain previous 
calibration data, press 

[>T<]

 (“Abort” will be 

displayed) then 

[FIX]

 at this time (or turn the scale 

off at any time during the process). 

 

Содержание RE series

Страница 1: ...RE Series Price Computing Scale RE C with Tower Model RE Model 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Страница 2: ......

Страница 3: ...Manual Pushbutton Tare 7 4 5 Pre set Keyboard Tare 7 4 6 Weighing and Pricing 8 4 7 Hold Function 8 4 8 Pre pack Weighing Function 8 4 9 PLU Function Price Lookup 9 4 10 By Count Function 10 4 11 Accu...

Страница 4: ...ed accessories and peripherals as available Operate the scale only under ambient conditions specified in these instructions Service should be performed by authorized personnel only 2 INSTALLATION 2 1...

Страница 5: ...AC Power Verify that the intended AC power source matches the AC adapter rating Connect the adapter to the power receptacle underneath and to the right of the scale then plug the AC adapter into a pro...

Страница 6: ...attery Power Saving Feature When powered on batteries the scale will enter sleep mode Weight display will show SLEEP and Unit Price and Total Price displays will be blank after the pre set minutes of...

Страница 7: ...D FUNCTIONS 3 1 Control Panel 3 2 Customer Display 1 Weight display 2 Unit Price display 3 Total Price display 4 Weights Measures marking 5 Leveling Indicator 6 Memory Keys Pull out Tab 7 Numeric Keyp...

Страница 8: ...ulation memory Activates the By Count operation and calculates the price of the item Stores or recalls a PLU In SETUP Mode ends the setup mode Holds the displayed Total Price when the item is removed...

Страница 9: ...15 seconds of inactivity and reactivates with a key press or item placed on the pan 4 3 Zero Operation Press 0 to zero the displayed weight the weight must be stable and within 2 of the scale capacit...

Страница 10: ...ice display The Unit Price and Total Price values are cleared automatically when the item is removed from the pan Unit Price will be retained if the scale is in Pre pack Mode see Section 4 8 4 7 Hold...

Страница 11: ...2 Enter the PLU number with the pre stored data then press PLU The corresponding Unit Price Tare value will be displayed 4 9 1 3 Recalling a PLU with auto PLU search function off Sec 5 2 Press PLU the...

Страница 12: ...tion is used as reference to calculate the price of non weighing items e g the price for 36 pieces when 3 pieces cost a total of 1 50 1 Enter the deal quantity 1 99 e g 3 then press X The X annunciato...

Страница 13: ...mode accordingly press 1 2 successively to price in per lb 1 2 annunciator will turn on per lb 1 4 annunciator will turn on return to normal Unit Price mode To enable Fractional Pricing in the Servic...

Страница 14: ...2 Sleep mode after 5 min Auto off disabled 3 Sleep mode after 5 min Auto off after 30 min 4 Sleep mode after 15 min Auto off disabled 5 Sleep mode after 15 min Auto off after 30 min Step 3 Automatic c...

Страница 15: ...s the switch inside the access hole A narrow screwdriver may be inserted through the hole to press the switch Note When accessing the bottom of the scale avoid placing the scale upside down on the pan...

Страница 16: ...o Ensure the pan is empty and the scale stable Press FIX to initiate countdown sequence 5 0 Countdown The countdown will repeat if the scale is not stable during this process Add20 2 3 calibration poi...

Страница 17: ...ving authorities the scale may be sealed in the following manner 6 2 1 Wire Sealing 1 Install the 2 cross drilled sealing screws securely as illustrated below Run the wire into the holes then apply th...

Страница 18: ...rgent DO NOT use acids alkalis strong solvents or abrasive cleaning agents Inspect the scale regularly Ensure that the housing pan and sub platform are clean and free from foreign material DO NOT allo...

Страница 19: ...roper calibration weights Weight display Weight exceeds power on zero range Remove excess weight from the pan during power on Weight display Overload condition Reduce the load on the pan Total Price d...

Страница 20: ...e Accumulation Keyboard Numeric 32 key tactile membrane switch with key press beeper Span Calibration 2 3 and full scale capacity Tare Range To full capacity by subtraction Zero Range At power on 10 k...

Страница 21: ...RE Series EN 19 8 2 Drawings Memory Keys Pull out Tab Template 13 6 345 mm 13 5 342 mm 17 432 mm 4 5 115 mm 16 7 423 mm 15 2 385 mm 9 5 242 mm 9 5 242 mm 14 356 mm 14 5 342 mm 13 6 345 mm 4 5 115 mm...

Страница 22: ...lated Thank you for your contribution to environmental protection For disposal instructions in Europe refer to www ohaus com weee FCC Note This equipment has been tested and found to comply with the l...

Страница 23: ...n damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than Ohau...

Страница 24: ...80252577 P N 80252577 2008 Ohaus Corporation all rights reserved Printed in China Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...

Отзывы: