EN 30 Ranger 4000 Count Series
7.5 Output Format
Weight string print format:
Field
Weight space Unit
space Stability space G/N
Space Term.
Char(s)
Length 9
1
5
1
1
1
NET
1
Each field is followed by a single delimiting space (ASCII: 32)
Definitions:
Weight
- Up to 9 characters, right justified, - at immediate left of most significant character
(if negative).
Unit
- Up to 5 characters, left justified. If the Unit in the Print Content menu was set to
OFF, the unit will be removed in the weight string and replaced by spaces.
Stability
-
e.
space is printed.
G/N
-
weight.
Terminating Character(s) - terminating character(s) printed depending on FEED menu setting.