To clear error press the standby button
Problem
Cause
Solution
Unit fails to power on
Missing or blown fuse
Add or replace fuse as necessary.
If problem persists, please contact your Ohaus representative for repair.
E1
Faulty temperature sensor
This error cannot be fixed by the end user.
Please contact your Ohaus representative for repair.
E2
Thermocouple failure or
Heating element failure
This error cannot be fixed by the end user.
Please contact your Ohaus representative for repair.
E3
Unit cannot reach set-point or
Probe not in thermometer well
If using probe, verify probe is in thermo well and follow Single Point Calibration
instructions on page 11. If problem persists, please contact your Ohaus representative
for repair.
Errors will cause heating function to cease. Timing functions will be unaffected.
15
t
roublesHooting