Valor 2000 Series Scales

The display flashes the span calibration point. Place the 
specified weight on the scale and press the 

Yes

 button.

 
To choose a different span point, repeatedly press the 

No

 

button to increment the selections.  When the desired 
value is displayed, place the specified weight on the scale 
and press the 

Yes

 button.

 
The display shows --C-- while the span point is 
established.
 
If span calibration was successful, dONE is displayed 
and the scale exits to the active weighing mode and 
displays the actual weight value.

 

 

 

 

 

 
 

 

 

4.2.2 Linearity Calibration 

Linearity calibration uses 3 calibration points.  The first calibration point is 
established with no weight on the scale.  The second calibration point is established 
at approximately half capacity.  The third calibration point is established at capacity.  
The Linearity calibration points are fixed and cannot be altered by the user during the 
calibration procedure.

 

When LIN is displayed, 

press the 

Yes

 button to access the 

Linearity Calibration menu item.
 
The display flashes 0.
 
With no weight on the scale, press the 

Yes 

button to 

establish the zero point.
 
The display shows --C-- while the zero point is 
established.

 

The display flashes the mid calibration point. Place the 
specified weight on the scale and press the 

Yes

 button.

 

The display shows --C-- while the mid point is 
established.
 

 

 

 

 
 

 

 

 

 

 

 

Содержание CS 2000

Страница 1: ...Valor 2000 Series TM Instruction Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Страница 2: ......

Страница 3: ...hese safety precautions Verify that the AC Adapter input voltage matches the local AC power supply Do not immerse the scale in water or other liquids Do not operate the scale in hostile environments D...

Страница 4: ...s Accepts the current setting on the display No Advances to the next menu or menu item Rejects the current setting on the display and advances to the next available setting 2 INSTALLATION 2 1 Package...

Страница 5: ...rear leveling feet must be screwed in tightly to avoid entry of water inside the scale 2 4 Power 2 4 1 AC Power The AC adapter included may be used to power the scale when battery power is not availab...

Страница 6: ...Fully Charged Flashing Green 70 100 Flashing Yellow 40 70 Flashing Red 0 40 The scale can operate for approximately 60 hours on a fully charged battery During battery operation the battery Indicator d...

Страница 7: ...ht display The scale must be stable to accept zero operation 3 3 Manual Tare Place a container on the weighing pan then press ON 0 T Off The display will show a net value of 0 To clear the Tare value...

Страница 8: ...ion power off the scale TABLE 4 1 CALIBRATION WEIGHTS Model Span Calibration weight kg User selectable Linearity Calibration weight kg Fixed V21PW3 1 2 3 2 3 V21PW6 2 4 6 4 6 V21PW15 5 10 15 10 15 4 2...

Страница 9: ...ity calibration uses 3 calibration points The first calibration point is established with no weight on the scale The second calibration point is established at approximately half capacity The third ca...

Страница 10: ...t kilograms ON or OFF When kg is lit press the Yes button to access the ON or OFF setting then press No to choose ON or OFF then Yes to accept the setting GRAM UNIT Set grams ON or OFF When g is lit p...

Страница 11: ...Menu LEGAL FOR TRADE LFT MODE Set the legal for trade status to ON or OFF 4 9 END End menu item 5 LEGAL FOR TRADE 5 1 Settings Enter the menu to verify the settings and perform a calibration as explai...

Страница 12: ...urchaser to ensure that all pertinent legal requirements are met Please contact your local weights and measures office for further details 5 3 Sealing The weights and measures official can apply a wir...

Страница 13: ...Move scale to suitable location Relieve internal pressure by loosening rear leveling feet Unable to calibrate LFT Menu set to ON Unstable environment Incorrect calibration mass Set LFT menu to OFF see...

Страница 14: ...g 3000 g x 0 5 g 6 lb x 0 001 lb 96 oz x 0 02 oz 6 kg x 0 001 kg 6000 g x 1 g 15 lb x 0 002 lb 240 oz x 0 05 oz 15 kg x 0 002 kg 15000 g x 2 g 30 lb x 0 005 lb 480 oz x 0 1 oz Maximum displayed resolu...

Страница 15: ...EN 13 Valor 2000 Series Scales 7 2 Drawings and Dimensions 249 mm 9 8 in 229 mm 9 0 in 233 mm 9 2 in 184 mm 7 2 in 142 mm 5 6 in Figure 7 1 Valor 2000 Series Overall Dimensions...

Страница 16: ...escriptive plate They may be put into service immediately Weighing Instruments to be verified in two stages have no green M metrology on the descriptive plate and bear one of the preceding identificat...

Страница 17: ...e radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will n...

Страница 18: ...to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than Ohaus In lieu of a properly returned warrant...

Отзывы: