EN-7
Valor 2000 Series Scales
The display flashes the span calibration point. Place the
specified weight on the scale and press the
Yes
button.
To choose a different span point, repeatedly press the
No
button to increment the selections. When the desired
value is displayed, place the specified weight on the scale
and press the
Yes
button.
The display shows --C-- while the span point is
established.
If span calibration was successful, dONE is displayed
and the scale exits to the active weighing mode and
displays the actual weight value.
4.2.2 Linearity Calibration
Linearity calibration uses 3 calibration points. The first calibration point is
established with no weight on the scale. The second calibration point is established
at approximately half capacity. The third calibration point is established at capacity.
The Linearity calibration points are fixed and cannot be altered by the user during the
calibration procedure.
When LIN is displayed,
press the
Yes
button to access the
Linearity Calibration menu item.
The display flashes 0.
With no weight on the scale, press the
Yes
button to
establish the zero point.
The display shows --C-- while the zero point is
established.
The display flashes the mid calibration point. Place the
specified weight on the scale and press the
Yes
button.
The display shows --C-- while the mid point is
established.