EN-4 Valor 2000
Series
Scales
2.4.2 Battery Power
The scale can be operated on the internal rechargeable battery when AC power is not
available. The scale will automatically switch to battery operation if there is a power
failure or the power cord is removed.
Before using the scale for the first time, the battery should be fully
charged for up to 12 hours. During charging, the battery Indicator
displays the charge level (see Table 2-1), and the scale can be operated
during charging. The battery is protected against over charging and the
scale can remain connected to the AC power line.
TABLE 2-1. BATTERY CHARGE STATUS.
Charge Status Indication Light
Charge level
Off
Fully Charged
Flashing Green
70 – 100 %
Flashing Yellow
40 – 70 %
Flashing Red
0 – 40 %
The scale can operate for approximately 60 hours on a fully charged battery.
During battery operation, the battery Indicator displays the battery status (see Table
2-2).
TABLE 2-2. BATTERY CAPACITY STATUS.
Charge Status Indication Light
Capacity Remaining
Solid Green
70 – 100 %
Solid Yellow
40 – 70 %
Solid Red
10 – 40 %
Flashing Red
0 – 10 %
Flashing Red, Lo bAt
Requires recharge;
Accuracy not guaranteed
NOTE:
• Charging the scale must only be performed in a dry environment.
• Install the rubber seal securely to ensure the scale is protected when
operating in a wet environment.
CAUTION:
Battery is to be replaced only by an authorized Ohaus service
dealer.
Risk of explosion can occur if replaced with the wrong type or connected
improperly.
Dispose of the lead acid battery according to local laws and regulations.