EN-4 Valor 2000 

Series 

Scales

2.4.2 Battery Power  

The scale can be operated on the internal rechargeable battery when AC power is not 
available. The scale will automatically switch to battery operation if there is a power 
failure or the power cord is removed. 
 

Before using the scale for the first time, the battery should be fully 
charged for up to 12 hours. During charging, the battery Indicator 
displays the charge level (see Table 2-1), and the scale can be operated 
during charging. The battery is protected against over charging and the 
scale can remain connected to the AC power line. 

 

TABLE 2-1. BATTERY CHARGE STATUS.

Charge Status Indication Light 

Charge level 

Off

Fully Charged

Flashing Green

70 – 100 %

Flashing Yellow

40 – 70 %

Flashing Red

0 – 40 %

 
The scale can operate for approximately 60 hours on a fully charged battery. 
During battery operation, the battery Indicator displays the battery status (see Table 
2-2). 

TABLE 2-2. BATTERY CAPACITY STATUS.

Charge Status Indication Light 

Capacity Remaining 

Solid Green

70 – 100 %

Solid Yellow

40 – 70 %

Solid Red

10 – 40 %

Flashing Red

 0 – 10 %

Flashing Red, Lo bAt

Requires recharge; 
Accuracy not guaranteed

 
NOTE: 

•  Charging the scale must only be performed in a dry environment. 
•  Install the rubber seal securely to ensure the scale is protected when  
    operating in a wet environment. 

 

CAUTION:

 Battery is to be replaced only by an authorized Ohaus service 

dealer. 
Risk of explosion can occur if replaced with the wrong type or connected 
improperly. 
Dispose of the lead acid battery according to local laws and regulations. 
 

Содержание CS 2000

Страница 1: ...Valor 2000 Series TM Instruction Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Страница 2: ......

Страница 3: ...hese safety precautions Verify that the AC Adapter input voltage matches the local AC power supply Do not immerse the scale in water or other liquids Do not operate the scale in hostile environments D...

Страница 4: ...s Accepts the current setting on the display No Advances to the next menu or menu item Rejects the current setting on the display and advances to the next available setting 2 INSTALLATION 2 1 Package...

Страница 5: ...rear leveling feet must be screwed in tightly to avoid entry of water inside the scale 2 4 Power 2 4 1 AC Power The AC adapter included may be used to power the scale when battery power is not availab...

Страница 6: ...Fully Charged Flashing Green 70 100 Flashing Yellow 40 70 Flashing Red 0 40 The scale can operate for approximately 60 hours on a fully charged battery During battery operation the battery Indicator d...

Страница 7: ...ht display The scale must be stable to accept zero operation 3 3 Manual Tare Place a container on the weighing pan then press ON 0 T Off The display will show a net value of 0 To clear the Tare value...

Страница 8: ...ion power off the scale TABLE 4 1 CALIBRATION WEIGHTS Model Span Calibration weight kg User selectable Linearity Calibration weight kg Fixed V21PW3 1 2 3 2 3 V21PW6 2 4 6 4 6 V21PW15 5 10 15 10 15 4 2...

Страница 9: ...ity calibration uses 3 calibration points The first calibration point is established with no weight on the scale The second calibration point is established at approximately half capacity The third ca...

Страница 10: ...t kilograms ON or OFF When kg is lit press the Yes button to access the ON or OFF setting then press No to choose ON or OFF then Yes to accept the setting GRAM UNIT Set grams ON or OFF When g is lit p...

Страница 11: ...Menu LEGAL FOR TRADE LFT MODE Set the legal for trade status to ON or OFF 4 9 END End menu item 5 LEGAL FOR TRADE 5 1 Settings Enter the menu to verify the settings and perform a calibration as explai...

Страница 12: ...urchaser to ensure that all pertinent legal requirements are met Please contact your local weights and measures office for further details 5 3 Sealing The weights and measures official can apply a wir...

Страница 13: ...Move scale to suitable location Relieve internal pressure by loosening rear leveling feet Unable to calibrate LFT Menu set to ON Unstable environment Incorrect calibration mass Set LFT menu to OFF see...

Страница 14: ...g 3000 g x 0 5 g 6 lb x 0 001 lb 96 oz x 0 02 oz 6 kg x 0 001 kg 6000 g x 1 g 15 lb x 0 002 lb 240 oz x 0 05 oz 15 kg x 0 002 kg 15000 g x 2 g 30 lb x 0 005 lb 480 oz x 0 1 oz Maximum displayed resolu...

Страница 15: ...EN 13 Valor 2000 Series Scales 7 2 Drawings and Dimensions 249 mm 9 8 in 229 mm 9 0 in 233 mm 9 2 in 184 mm 7 2 in 142 mm 5 6 in Figure 7 1 Valor 2000 Series Overall Dimensions...

Страница 16: ...escriptive plate They may be put into service immediately Weighing Instruments to be verified in two stages have no green M metrology on the descriptive plate and bear one of the preceding identificat...

Страница 17: ...e radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will n...

Страница 18: ...to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than Ohaus In lieu of a properly returned warrant...

Отзывы: