33
Press scroll key until the display shows . Send a 100 ohms signal to the RTD input terminals
according to the connection shown below:
Step 4.
18
19
20
100 ohms
Press scroll key for at least 3 seconds. The display will blink a moment, otherwise the calibration fails.
Figure 5.1 RTD Calibration
Press scroll key and the display will show . Change the ohm's value to 300 ohms .Press scroll key
for at least 3 seconds. The display will blink a moment and two values are obtained for RTDH and RTDL
(step 4). Otherwise, if the display didn't blink or if any value obtained for RTDH and RTDL is equal to
-199.9 or 199.9 , then the calibration fails.
Step 5.
Perform step 6 to calibrate offset of cold junction compensation, if required.
Setup the equipments according to the following diagram for calibrating the cold junction compensation.
Note that a K type thermocouple must be used.
Step 6.
*
ETR-4100
ETR-8100
Stay at least 20 minutes in still-
air room
room temperature 25 ± 3 °C
Stay at least 20 minutes in still-
air room
room temperature 25 ± 3 °C
Figure 5.2
Cold Junction Calibration Setup
Figure 5.2
Cold Junction Calibration Setup
19
20
K+
K
5520A
Calibrator
5520A
Calibrator
K-TC
The 5520A calibrator is configured as K type thermocouple output with internal compensation. Send a 0.00°C signal
to the unit under calibration.
Perform step 7 to calibrate gain of cold junction compensation, if required.
Setup the equipments like step 6. The unit under calibration is powered in a still-air room with
temperature 50 ±3°C. Wait at least 20 minutes for warming up . The calibrator source is set at 0.00°C
with internal compensation mode.
Step 7.
The unit under calibration is powered in a still-air room with temperature 25±3 °C. Stay at least 20 minutes
for warming up. Perform step 1 stated above, then press scroll key until the display shows . Press
up/down key to obtain 40.00. Press scroll key for at least 3 seconds. The display will blink a moment and a
new value is obtained. Otherwise, if the display didn't blink or if the obtained value is equal to -5.00 or
40.00, then the calibration fails.
*
ETR-4100
ETR-8100
ETR-4100
ETR-8100
4
5
6
ETR-9100
5
6
ETR-9100