8
23.
Make sure the filter is in place inside the receiver bottom cap.
24.
Pour water into the hole in the receiver bottom cap.
25.
Open the outlet valve to allow excess water to drain, then close the outlet
valve.
26. Apply grease to the receiver cap o-rings and place one around each cap.
27.
Screw the receiver top cap onto one end of the receiver body.
28.
Screw the receiver body onto the bottom receiver cap.
29.
Screw the valve stem into the port on the receiver top cap.
30.
Attach the CO
2
Pressure Assembly to the top valve stem and secure it
with a safety pin.
31.
Pull out the red knob on the heating jacket and lower the cell. Rotate the
cell until it seats.
32.
Move the thermometer to the cell body.
33.
Place a CO
2
bulb in the CO
2
pressure assembly. Set the pressure to the
appropriate back pressure for your test.
34.
Stroke the pump to apply the same pressure to the cell.
35.
When the cell has reached the test temperature, stroke the pump to apply
the working test pressure to the cell.
36. Collect filtrate by opening the needle valve on the LCM Receiver.
37.
At the end of the test, turn off the heating jacket.
38.
Allow the cell to cool.
39. Close the regulator on the CO
2
pressure assembly.
40. Open the pressure release valve on the pump to release pressure inside
the cell.
41.
Unscrew the VCO fitting to disconnect the LCM Receiver from the cell
cap.
42.
Disconnect the inlet pressure manifold from the inlet pressure assembly.
43.
Remove the cell from the heating jacket and set it upright to disassemble.