Connect the LCM Receiver bottom cap assembly (bottom cap, needle
valve, and female VCO fitting) to the male VCO fitting on the cell cap as-
sembly. Make sure the filter is in place inside the receiver bottom cap.
26. The space between the filter media and the receiver bottom cap should
be filled with water to ensure that the volume of filtrate passing through
the filter will displace an equal volume of filtrate into the receiver.
a.
Pour enough water into the hole in the LCM Receiver bottom cap to fill
the void space.
b.
Open the needle valve and allow the excess water to drain.
c.
Close the need valve.
27.
Apply a thin coat of grease to the receiver cap o-rings (#171-84-015).
28.
Place one o-ring around each cap.
29.
Screw the top cap into one end of the cell body (the ends of the body are
interchangeable).
30.
Screw the receiver body onto the bottom receiver cap.
31.
Screw a valve stem into the port in the top receiver cap.
32.
Attached the CO2 Pressure Assembly (#170-04) to the top valve stem
and secure it with the safety pin (#171-23-1).
33. Hold the outlet pressure assembly with one hand and pull the stop on
the heating jacket out of the way to lower the cell fully inside the heating
jacket.
Filter (#171-84-014)
O-ring (#171-84-015)