OfiTE 120-60 Скачать руководство пользователя страница 6

OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com

5

4.  Edit Setpoint 2 / Increase Setpoint Digit

Press this button to change the timer value of Setpoint 2. When 

editing a setpoint value, this button increases the second digit of 

the value by 1.

5.  Increase Setpoint Digit

This button increases the fourth digit of the setpoint value by 1.

6.  Increase Setpoint Digit

This button increases the third digit of the setpoint value by 1.

7.  Enter Button

Press this button to store the Setpoint value in memory.

8.  Timer Reset

Press this button to reset the timer to zero.

9.  Keypad: VAR

This button accesses the PID Parameter Control Menu.

10. Keypad: SETPT

Press this button to change the rotational speed of Setpoint 1 or 

Setpoint 2.

11. Keypad: ENTER

Press this button to store the Setpoint value in memory.

12. Two-Way Switch: ON - OFF 

This switch provides power to the unit.

13. Two-Way Switch: MAN/AUTO - Variable

This switch determines the operating mode of the unit. If it is set to 

MAN/AUTO, the MAN - AUTO switch (15) becomes active. If it is 

set to VARIABLE, the unit will be in Variable Mode.

Содержание 120-60

Страница 1: ...30 Volt 120 60 F For Fracturing Fluids 115 Volt 120 60 1 F For Fracturing Fluids 230 Volt Instruction Manual Updated 11 16 2015 Ver 5 0 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77...

Страница 2: ...320 7300 Fax 713 880 9886 www ofite com 1 Intro 2 Description 2 Specifications 3 Components 3 Installation 7 Operation 8 Manual Mode 8 Variable Mode 9 Auto Mode 9 Timer 10 Speed Controller 10 PID Para...

Страница 3: ...be utilized to mix cements according to the procedures stated by the API The proper amount of mix water is carefully weighed and poured into the mixing container of the blender The rotational speed is...

Страница 4: ...5 Volt Only Qty 4 122 200 Blending Assembly Square Drive 1 qt Qty 12 122 202 Stainless Steel Container 1 qt 122 203 Container Lid 1 qt 122 204 Bottom Gasket Qty 6 120 61 1 Spare Parts Kit 230 Volt 122...

Страница 5: ...display shows the speed of the blade On models for cement testing 120 60 120 60 1 multiply the value displayed here by 10 to determine the actual RPM of the blade 2 Timer Display Seconds 3 Edit Setpo...

Страница 6: ...point value by 1 7 Enter Button Press this button to store the Setpoint value in memory 8 Timer Reset Press this button to reset the timer to zero 9 Keypad VAR This button accesses the PID Parameter C...

Страница 7: ...For testing Cement part numbers 120 60 and 120 60 1 Setpoint 1 is 4 000 RPM and Setpoint 2 is 12 000 RPM For testing Fracturing Fluids Setpoint 1 is 500 and Setpoint 2 is 1 000 RPM 15 Three Way Switch...

Страница 8: ...fely away from sinks and other possible hazards 2 Place the blending assembly on the base of the unit 3 Plug the power supply cord into the port underneath the cabinet housing labeled BLENDER 4 Plug t...

Страница 9: ...nd turn the Potentiometer 16 fully counter clockwise 1 Place the mix water into the mixing cup and place the lid on top of the container 2 The MAN AUTO Variable switch 13 should be in the MAN AUTO pos...

Страница 10: ...er position Place the MAN AUTO VARIABLE switch 13 to the MAN AUTO position Place the MANUAL AUTO switch 15 switch to the AUTO position Once the switch is placed into the AUTO position the blender shou...

Страница 11: ...e display 3 6 each increase a separate digit of the setpoint value The digits will scroll from 0 to 9 and then return to 0 3 When you reach the correct setpoint value press the enter button 7 to store...

Страница 12: ...ut the response was slow increase the value of Variable 9 by 4 Repeat the above procedure changing the value of VARIABLE 9 until there is minimal overshoot and the ramp is smooth Note that to fine tun...

Страница 13: ...ny losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate hea...

Отзывы: