During a test, the computer will be continuously adjusting the pressure inside
the cell. As the pressure increases due to thermal expansion, the computer
will allow pressure to bleed. Likewise, if pressure begins to drop, the
computer will add pressure to keep it within the target range.
There is a secondary
Regulator
(#120-50-037) which controls the air
pressure to the pressure relief valve.
This regulator comes calibrated and requires no adjustment unless a part
within this system has been replaced.
If the user must gain access to the regulator, simply remove the back panel
of the instrument. The regulator and gauge are located inside. There is one
set on each side. To set the regulator to the proper pressure, apply pressure
to the system and begin opening the regulator between 50 - 80 psi (344.7 -
551.6 kPa). If the pressure relief valve does not open, continue adjusting the
regulator to a higher pressure as needed to obtain optimal operation of the
relief valve.
Maintenance
Pressure Relief Valve
Regulator
Air Regulator
(#120-50-037)
Gauge
(#143-01)