Description
9
GA12119_00_A3 - 01/2012 - © Oerlikon Leybold Vacuum
1.3
Technical Data
Cold head COOLPOWER
5/100
5/100
5/100 T
140 T
Part No.
893 04
893 05
129 78
842 030
Compressor unit COOLPAK
6000, 6200
6000, 6200
6000, 6200
6000, 6200
Refrigerating capacity at 50/60 Hz
1st stage at 80 K, approx.
100 W
100 W
100 W
140 W
1st stage at 20 K, approx.
–
–
–
20 W
2nd stage at 20 K, approx.
6 W
6 W
7 W
–
2nd stage at 10 K, approx.
–
–
3 W
–
Lowest achievable
temperature, approx.
1st stage
#
35 K
#
35 K
#
35 K
#
15 K
2nd stage
#
10 K
#
10 K
6 K
–
Cooldown time
of 2nd stage to 20 K, approx.
20 min
20 min
20 min
–
to 20 K, approx.
–
–
-
55 min
Permissible ambient temperature
10-40 °C
He filling pressure
at room temperature 1)
16(-1) bar
Opening pressure
of safety valve
28 bar
He connections:
Self-sealing fittings
1/2“, m 2)
Weight excluding power cord, approx.
8 kg
Connection for mounting
on the system
Flange
Weld-on pipe
Weld-on pipe
Flange
DN 160 ISO-K
[
130 mm
[
130 mm
DN 160
ISO-K
Holes with a thread M4
at the copper flanges
1st stage
16
16
16
8
2nd stage
6
6
6
–
Length of power cord
to the compressor unit
4.5 m
Type of protection
provided by the housing
DIN/VDE Standard 0470-1
IP 52
1)
State of delivery; for filling pressure during operation refer to the Operating Instructions for the compressor unit.
2)
f = female coupling, m = male coupling; Type „Aeroquip series 5400“ or compatible types