Beam Alignment Mask
BEAM ALIGNMENT VERIFICATION
1. Load a film cassette in the alignment tool. Make the following four exposures on one sheet of film:
2. Select
Manual - Fluoro, Normal field
and then
Film
mode. Take a 50 kVp @ 0.9 mAs exposure.
3. Select
Manual - Fluoro, MAG 1 field
and then
Film
mode. Take a 50 kVp @ 0.9 mAs exposure.
4. Select
Manual - Fluoro, MAG 2 field
and then
Film
mode. Take a 50 kVp @ 0.9 mAs exposure.