9600 Periodic Maintenance Procedure
00-880606-02
Page 19
ELECTRICAL PLUG/POWER CORD
1. Inspect the Workstation AC power cable for damage and wear.
INTERCONNECT CABLE and PINS
1. Inspect the interconnect cable; verify that the cable, Mobile C-Arm connector, and pins are not
damaged. Verify the LEMO connector strain relief is torqued to 60 inch/lbs.
MOTORIZED FUNCTIONS / LIMIT SWITCHES
WARNING
Observe safety procedures when operating motorized features.
1. Make sure the Interconnect Cable is connected to the C-Arm.
2. Plug the AC power cord into the wall receptacle.
3. Turn the Workstation keyswitch to
ON
.
VERTICAL LIFT
1. Raise the vertical column until the upper limit switch is activated.
2. Lower the vertical column until the lower limit switch is activated.
3. Verify smooth motor operation.
4. Observe that no drift occurs when the switch is released.
L-ARM ROTATION
NOTE:
This test is NA (non-applicable) if testing a Super-C system (which does not use an L-Arm).
1. Fully retract the cross arm.
2. With the L-Arm in the lower vertical position, rotate the L-Arm approximately 190 degrees counter
clockwise, until the limit switch engages and disables motor operation.
3. Rotate the L-Arm back to the lower vertical position and then rotate the L-Arm approximately 190
degrees clockwise, until the limit switch engages and disables motor operation. Motor operation
should be smooth with no vibration.