9600 Periodic Maintenance Procedure
00-880606-02
Page 32
4. Boost is automatically enabled and the Boost Enable LED on the C-Arm control panel illuminates.
5. Boost mode cannot be disabled.
6. The
PULSE
and
BOOST ENABLE
keys on the C-Arm control panel are disabled; the Cardiac
software blocks all input from these keys.
7. Chole and One-Shot modes are disabled.
8. Averaging rates are 1, 2, 4, or 8 FPS (frames-per-sec.); the default rate is 1 FPS.
9. The only Acquire Image rate allowed is 30 FPS (for 60 Hz systems) or 25 FPS (for 50 Hz systems).
10. Pressing
BOOST
activates Pulse mode.
11. Pulse mode generates a fixed rate of 30 PPS (pulses-per-sec.) for 60 Hz systems, or 25 PPS for 50
Hz systems. Other pulse rates cannot be selected and the Select Pulse Rates screen is not available.
12.
SCOUT FLUORO
is not active in pulse mode.
13. Maximum pulse (Boost) exposure rate is 20 seconds; after 20 sec., exposure is automatically
terminated and BOOST OVERTIME displays on the control panel.
14. Maximum mA is 60 mA.
15. kV range is 40 - 120 kV.
LAST IMAGE HOLD
Verify proper operation of Last Image Hold by using Recall (Not on SP):
1. Save a few images and note the order they were saved in.
2. Press
the
RECALL
key.
4. Verify the last image saved is displayed on the left monitor.
COOLING FAN (OPTION)
Applicable only for systems equipped with the 9600 Cooling Kit (Std. C-Arm and Super C-Arm) which
includes the ventilated HV cable cover.
1. Switch
OFF
the Workstation keyswitch, then remove the HV cable cover from the X-ray tube but do
not disconnect the cooling fan wires.
2. Switch
ON
the Workstation keyswitch and ensure that the fan turns on at approx. the same time the
keyswitch is switched on.
3. Ensure proper fan operation. Blade rotation should not be sluggish and there should be no
obstructions inside the fan itself. Fan air should blow air directly onto the X-ray tube heat sink.