Installation procedure | 9
Installation procedure
The procedure described here assumes that ThermoServer/
ThermoClient software is installed on your computer, and that your
Oceasoft wireless receiver is configured and ready to use.
Before you begin
1. Connect the temperature probe firmly to one end of the capillary
cable by joining the connectors (without unscrewing them).
2. Connect the other end to the Cobalt 2 wireless module.
Placing your Cobalt 2 module for best performance
For optimal operation, follow these recommendations when physically
placing your device:
Do not place the module within 40 cm (16 inches) of another
module.
Make sure the wireless module is not placed on an electrical
conduit or cable tray (such as for computer network cables).
For best results, place the module so that it faces the general
direction of the receiver antenna. Modules without displays may
be placed on top of refrigerators or freezers.
Keep around 20 cm (8 inches) clear space around the module.
A module that is “stuck” between two refrigerators may not
communicate effectively.
Make sure all cables, if any, are firmly attached, that you can
open and close the enclosure door without forcing, and that the
capillary cable remains flush with the enclosure door joint.
See Appendix (
pg 17
) for additional considerations.
Placing the probe within the enclosure
Depending on your refrigerator or freezer model, it may be better to
leave the probe sitting on a shelf inside the enclosure rather than
sticking it to a side wall. This is particularly relevant in cases where you
have carried out a mapping study and determined the ideal location for
the probe. In that case, make sure that the probe does not interfere with
loading and unloading of your product(s) and that the location does not
represent a risk either for your product(s) or the probe.