TROUBLESHOOTING

TROUBLESHOOTING

1. 

Unplug power cord.

2. 

The foot control is connected

to a junction box adjacent to the

control module located on the top

end of the Ultrasound table.

3. 

Remove the locking clip on the plug.

4. 

Pull firmly to remove the plug.

5. 

Obtain the new foot control and 

insert the cable plug into the junction box until firmly seated.

6. 

Replace the locking clip.

7. 

Reconnect power cord.

8. 

Attempt to operate the table elevation functions from the foot control.

9. 

Should this fail to return the Ultrasound table to an operating condition,  

 

contact the customer service department.

   FOOT CONTROL REPLACEMENT INSTRUCTIONS

   HAND CONTROL REPLACEMENT INSTRUCTIONS

Replacement of the foot control may be necessary if the foot control does not actuate 

the functions of elevation and the hand control is operational.  See troubleshooting 

section for details.
To replace the foot control, follow these steps:

IMPORTANT

Always unplug power 

cord before removing or 

connecting cables to the 

control box.

SAFETY NOTE:

1. 

Unplug power cord.

2. 

Release the locking clip on
control box.

3. 

Remove the plug from the 
control box socket by pulling
firmly outward.

4. 

Obtain new hand control.

5. 

Insert the plug into the socket, applying pressure until it is firmly seated.

6. 

Replace the locking clip on the control box.

7. 

Reconnect power cord.

8. 

Operate hand control and determine if proper hand control function 
has returned.

9. 

Should this fail to return the Ultrasound table to an operating condition,  

 

contact the customer service department.

Locking Clip

Locking Clip

Replacement of the hand control may be necessary if the hand control does not actuate 

the functions of elevation and the foot control is operational.  See troubleshooting 

section for details.
To replace the foot control, follow these steps:

Содержание WOMEN'S IMAGING

Страница 1: ...les I N S T R U C T I O N M A N U A L www oakworksmed com 717 235 6807 GENERAL WITH FOWLER WOMEN S IMAGING GENERAL MULTI SPECIALTY GENERAL 3 SECTION ECHOCARDIOGRAPHY VASCULAR VASCULAR WITH FOWLER made...

Страница 2: ...ithout notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc assumes no responsibility for any errors that may appear in this document nor does it make expressed or implied w...

Страница 3: ...owler 14 Handset Safety Lock Out 15 Side Rail Use 15 Retractable Foot Rest stirrup Use 15 Adjustable Angle Back Support 16 Central Base Locking System 16 Locking Casters 16 Operating the Manual Adjust...

Страница 4: ...n is given regarding the recommended temperature limits during transport and storing This symbol when used in this manual or on product labels indicates the date of manufacture of the device This symb...

Страница 5: ...ent The patient may lose balance and fall This table is not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide To reduce the risk of burns fire electr...

Страница 6: ...SE 15 PADDING 3 8 cm multi layer UPHOLSTERY TerraTouch over 20 colors TABLE WEIGHT 313 lbs 142kg WARRANTY 3 years parts 1 year labor CERTIFICATIONS FDA listed cETL approved CE marked INCLUDED FEATURES...

Страница 7: ...BURG REVERSE 15 PADDING 3 8 cm multi layer UPHOLSTERY TerraTouch over 20 colors TABLE WEIGHT 313 lbs 142kg WARRANTY 3 years parts 1 year labor CERTIFICATIONS FDA listed cETL approved CE marked INCLUDE...

Страница 8: ...ed PRODUCT SPECS INCLUDED FEATURES Electric Height Range 22 38 Width Specify 30 or 27 Trendelenburg Reverse Tilt 15 Electric Fowler Central Base Locking System Hand Control Height Trendelenburg Revers...

Страница 9: ...t Right Sliding Cardio Panel Left side Manual Locking Casters Paper Roll Holder OPTIONS ACCESSORIES OPTIONS Central Base Locking System Side rails 2 Sliding Cardio Panel Electric Upgrade 30 w Left sid...

Страница 10: ...l motions Paper Roll Holder OPTIONS ACCESSORIES OPTIONS Central Base Locking System Sliding Cardio Panel Manual 30 w Left side only Adjustable Access Back Support Left Adjustable Access Back Support R...

Страница 11: ...cm TRENDELENBURG 15 REVERSE TRENDELENBURG 40 PADDING 3 8 cm multi layer UPHOLSTERY TerraTouch over 20 colors TABLE WEIGHT 313 lbs 142kg WARRANTY 3 years parts 1 year labor CERTIFICATIONS FDA listed c...

Страница 12: ...circuit for the United States and a nominal 220 volt circuit for use in Europe The plug has a grounding pin in a pattern that varies by country as shown in the Product Identification section of this m...

Страница 13: ...op tilt is preserved during elevation Table DOWN Control lowers tabletop table is lowered from any position CAUTION Do not sit on the elevated Back Rest or Leg Rest CAUTION Do not leave clients unatte...

Страница 14: ...ilt is preserved during elevation Table DOWN Control lowers tabletop table is lowered from any position Foot End Trendelenburg Control activates tilt up to 15 positioning stops when table is level pre...

Страница 15: ...t up to 15 positioning stops when table is level press again for complete reverse Foot End Backrest activates backrest up to 75 Backrest activates backrest down to level position B k t Cardio Panel CL...

Страница 16: ...rendelenburg Control activates tilt up to 15 positioning stops when table is level press again for complete reverse Foot End Trendelenburg Control activates tilt up to 15 positioning stops when table...

Страница 17: ...ntrol lowers tabletop table is lowered from any position Foot End Trendelenburg Control activates tilt up to 15 positioning stops when table is level press again for complete reverse Head End Trendele...

Страница 18: ...plying light outward pressure RETRACTABLE FOOT REST STIRRUP USE Fully lower the manual adjustable foot rest Fold out to fully open position Adjust length width of stirrups as needed Repeat for second...

Страница 19: ...lowered to desired position CAUTION Do not sit on Adjustable Angle Back Support Max Capacity 150 lbs TO LOCK Press down on locking arm with your foot to lock table to the floor TO UN LOCK Lift up on...

Страница 20: ...E THE LEG SECTION Raise the leg section by pulling it up from the lower center end of the table to the proper position for the procedure Use a slow or medium speed If you raise the leg section too fas...

Страница 21: ...with one hand at 90 bend and rotate foot board 180 with second hand until in the up position Grasp with one hand at 90 bend and pull away from table to lock in up position Check locking by moving foot...

Страница 22: ...board and rotate 180 to up position Grasp rod at 90 bend and pull away from table to lock in position CAUTION When the footboards are in the down position it is possible for them to hit the ground in...

Страница 23: ...strap under rail attached to the top NOTE For Vascular tables the strap is installed under the rail attached to the top and above the second rail FLAT TOP TABLES Install strap under backrest above al...

Страница 24: ...Mounting Plate PAPER HOLDER USE To load the paper pull down on two paper holder support arms located on the under side of backrest Rotate to 90 position Insert rod thru paper roll and place rod into...

Страница 25: ...ree When replacing the QuickLock Face Rest just push the frame back into the outlets being sure to keep the dowels within 3 of the table to ensure a stable connection For complete use care instruction...

Страница 26: ...ail accessory Attach or remove using knobs DETACHABLE ARM BOARD This product can be used on any of Oakworks ultrasound tables with a T Rail Adapter and can support a variety of positioning configurati...

Страница 27: ...e hand control is operational See troubleshooting section for details To replace the foot control follow these steps IMPORTANT Always unplug power cord before removing or connecting cables to the cont...

Страница 28: ...eral attempts to press them simultaneously b Control box will beep approximately 10 times c Run all motors in system to fully retracted position down d Re test table to insure proper functions Unplug...

Страница 29: ...be sure to unplug the power cord from the power outlet INSPECTIONS MAINTENANCE RECOMMENDED REGULAR INSPECTIONS monthly or local standard Check for damage to the power handset or footswitch cables Visu...

Страница 30: ...cular Motion Actuator 61286 Paper Roll Holder Rod 61736 06 Paper Hold Down Strap 64327 Strap Vascular Patient Restraint 64432 4 Pedal Foot Control 64434 6 Pedal Foot Control 64431 8 Pedal Foot Control...

Страница 31: ...AC 15amp 60Hz 4 0 amps 6 0 amps 220 VAC 10amps 50Hz 2 0 amps 3 0 amps 100 VAC 20amp 50 60 Hz 4 8 amps 7 2 amps 24 VDC Class I Equipment Type B Applied Part IPX0 Continuous Operation 10 Duty Cycle Weig...

Страница 32: ...Ultrasound Table uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 1...

Страница 33: ...rential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage va...

Страница 34: ...mined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1...

Страница 35: ...32 THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 36: ...Inc 923 East Wellspring Road New Freedom PA 17349 Toll Free USA only 800 916 4613 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com OAKWORKS Dual Tower Ultrasound Tables European Compliance for...

Отзывы: