CLEANING & DISINFECTION / INSPECTIONS / WARRANTY

CLEANING PROCESS

To reduce the risk of electric shock:

-  Always unplug this furnishing form the electrical outlet before cleaning.

DANGER

INSPECTIONS

WARRANTY

Follow the cleaners/disinfectant manufacturers’ directions for use. Please note that cleaning and disinfecting the  

PX150 & PX200 Beds is a two part process. First it must be cleaned of any visible soil, then it can be disinfected. The bed should be 

cleaned as soon as possible after use. Please follow these procedures for best results:

1.  Using a fresh wipe or wipes, clean any visible soil off of the bed, working from the top to the bottom of the bed.
2.  Throw the used wipe(s) away.
3.  Using a fresh wipe or wipes, thoroughly wipe all surfaces of the top and any high-contact areas such as handles, 

handsets, etc., making sure they remain wet for the disinfectant manufacturer’s recommended contact time.

4.  Throw the used wipe(s) away and allow the bed to dry.

RECOMMENDED REGULAR INSPECTIONS (monthly or local standard)

• Visually inspect components for obvious damage that could cause problems during operation.

RECOMMENDED PERIODIC INSPECTIONS (yearly or local standard)

• Visually inspect components for obvious damage that could cause problems during operation.
• Check that all fasteners are present and fastened securely.
• Clean unusual buildup of dirt on the bed and/or parts of the bed not normally cleaned on a regular basis.
• Check for tears or cracks in the coverings.

3 Years Parts, 2 Years Labor

Содержание PX150

Страница 1: ...USER MANUAL www oakworksmed com 717 235 6807 made in the USA with US imported parts PX150 PX200 EMERGENCY FIELD BEDS...

Страница 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifica tions and the like documen...

Страница 3: ...Back Rest 8 Side End Rail Use 8 Moving the bed 9 Dual Locking Casters Use 9 Cleaning Disinfection Recommended Cleaners Disinfectants 10 Cleaning Process 11 Inspections Maintenance 11 Warranty Informat...

Страница 4: ...al or other medical facility The bed is only to be moved within the patient room for cleaning or patient access No spe cial training is required but a review of the following manual is important for t...

Страница 5: ...injury to persons 1 Unplug this furnishing from the electrical outlet before cleaning 2 Unplug from outlet before adding or removing parts 3 Never operate this furnishing if it has a damaged cord or...

Страница 6: ...Otherwise individuals may become trapped between the bed rail and the bed frame Bed rails can be deformed or broken if excessive side pressure is exerted on the bed rails These bed rails are used for...

Страница 7: ...st Mattress 6 Thick 15 cm Removable Water Resistant Foot Control Height Motion Bed Voltage Options 120V 60Hz or 230V 50Hz Bed Weight 180 lbs 82 kg Bed Capacity 400 lbs 182 kg patient and accessories W...

Страница 8: ...rest Width 36 91 cm Bed Weight 200 lbs 91 kg 4 Button Hand Control Height Backrest Padding 6 Thick 15 cm Removable Water Resistant Bed Voltage Options 120V 60Hz or 230V 50Hz Bed Capacity 400 lbs 182 k...

Страница 9: ...plugged into an appropriate outlet that is prop erly installed and grounded in accordance with all local codes and ordinances See U S sample below Correct Implementation Incorrect Implementation Risk...

Страница 10: ...CONTROL OPERATION DIRECTIONS FOR USE Height DOWN Height UP The Oakworks PX200 bed offers height and backrest positioning with electronic ease Operate the Hand Control as shown below to raise or lower...

Страница 11: ...e of their weight off the back rest while you lift CAUTION ADJUSTING THE BACK REST SECTION PX150 BEDS ONLY Do not sit on the Backrest section 8 DIRECTIONS FOR USE SIDE END RAIL USE Firmly grab the han...

Страница 12: ...to protect the surface from any debris from the bottom of your foot control Two people are required to safely move the bed Grasp the bed frame securely at the end with either the locking casters or t...

Страница 13: ...KWORKS recommends a prepackaged wipe for cleaners disinfectants to ensure best distribution of disinfectant for the required kill time without leaving excess residue and or overexposing components the...

Страница 14: ...he bed 2 Throw the used wipe s away 3 Using a fresh wipe or wipes thoroughly wipe all surfaces of the top and any high contact areas such as handles handsets etc making sure they remain wet for the di...

Страница 15: ...lways start with a letter The model number and serial number are located on the underside of the seat section Model Number Serial Number 16796025 PXEXFT311873ACTTVN Ser PXR563146 PX FT 30 x 22 38 x 72...

Страница 16: ...amp 50 Hz Current Draw 2 7 amps 1 5 amps 3 3 amps Maximum Momentary Current Consumption 9 0 amps 5 0 amps 11 0 amps Voltage Output to Actuators 24 VDC 24 VDC 24 VDC Electric Shock Protection Class 1...

Страница 17: ...c environment in which radiated RF disturbances are controlled the customer or the user of the bed can help prevent electromagnetic interference by maintaining a minimum distance between portable and...

Страница 18: ...al mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical com mercial or hospital environ ment Voltage dips short inter ruptions and voltage vari...

Страница 19: ...RF transmitters as deter mined by an electromagnetic site survey ashould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the...

Страница 20: ...reedom PA 17349 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com made in the USA with US imported parts 3034177 CONFORMS TO ANSI AAMI STD ES60601 1 IEC STD 60601 1 3RD EDITION IEC STD 60601 1 2...

Отзывы: