SPECIFICATIONS

Guidance and manufacturer’s declaration - electromagnetic immunity

The bed is intended for use in the electromagnetic environment specified below. The customer or the user 

of the bed should assure that it is used in such an environment.

Immunity Test

IEC 60601 test level

Compliance level

Electromagnetic

environment - guidance

Conducted RF 

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2,5 GHz

3Vrms

3V/m

Portable and mobile RF communications equipment should 

be used no closer to any part of the bed, including cables, 

than the recommended separation distance calculated from 

the equation applicable to the frequency of the transmitter.

Recommended separation distance

d = 1.2 √ P

 

d = 1.2 √ P

   80 MHz to 800 MHz

d = 2.3 √ P 

  800 MHz to 2,5 GHz

Where P is the maximum output power rating of the trans-

mitter in watts (W) according to the transmitter manufacturer 

and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as deter-mined 

by an electromagnetic site survey, 

a

should be less than the 

compliance level in each frequency range. 

b

Interference may occur in the vicinity of equipment marked 

with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

a

  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accu-

racy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should 

be considered. If the measured field strength in the location in which the bed is used exceeds the applicable RF com-

pliance level above, the bed should be observed to verify normal operation. If abnormal performance is observed, 

additional measures may be necessary, such as re-orienting or relocating the bed

b

  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.

Содержание PX150

Страница 1: ...USER MANUAL www oakworksmed com 717 235 6807 made in the USA with US imported parts PX150 PX200 EMERGENCY FIELD BEDS...

Страница 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifica tions and the like documen...

Страница 3: ...Back Rest 8 Side End Rail Use 8 Moving the bed 9 Dual Locking Casters Use 9 Cleaning Disinfection Recommended Cleaners Disinfectants 10 Cleaning Process 11 Inspections Maintenance 11 Warranty Informat...

Страница 4: ...al or other medical facility The bed is only to be moved within the patient room for cleaning or patient access No spe cial training is required but a review of the following manual is important for t...

Страница 5: ...injury to persons 1 Unplug this furnishing from the electrical outlet before cleaning 2 Unplug from outlet before adding or removing parts 3 Never operate this furnishing if it has a damaged cord or...

Страница 6: ...Otherwise individuals may become trapped between the bed rail and the bed frame Bed rails can be deformed or broken if excessive side pressure is exerted on the bed rails These bed rails are used for...

Страница 7: ...st Mattress 6 Thick 15 cm Removable Water Resistant Foot Control Height Motion Bed Voltage Options 120V 60Hz or 230V 50Hz Bed Weight 180 lbs 82 kg Bed Capacity 400 lbs 182 kg patient and accessories W...

Страница 8: ...rest Width 36 91 cm Bed Weight 200 lbs 91 kg 4 Button Hand Control Height Backrest Padding 6 Thick 15 cm Removable Water Resistant Bed Voltage Options 120V 60Hz or 230V 50Hz Bed Capacity 400 lbs 182 k...

Страница 9: ...plugged into an appropriate outlet that is prop erly installed and grounded in accordance with all local codes and ordinances See U S sample below Correct Implementation Incorrect Implementation Risk...

Страница 10: ...CONTROL OPERATION DIRECTIONS FOR USE Height DOWN Height UP The Oakworks PX200 bed offers height and backrest positioning with electronic ease Operate the Hand Control as shown below to raise or lower...

Страница 11: ...e of their weight off the back rest while you lift CAUTION ADJUSTING THE BACK REST SECTION PX150 BEDS ONLY Do not sit on the Backrest section 8 DIRECTIONS FOR USE SIDE END RAIL USE Firmly grab the han...

Страница 12: ...to protect the surface from any debris from the bottom of your foot control Two people are required to safely move the bed Grasp the bed frame securely at the end with either the locking casters or t...

Страница 13: ...KWORKS recommends a prepackaged wipe for cleaners disinfectants to ensure best distribution of disinfectant for the required kill time without leaving excess residue and or overexposing components the...

Страница 14: ...he bed 2 Throw the used wipe s away 3 Using a fresh wipe or wipes thoroughly wipe all surfaces of the top and any high contact areas such as handles handsets etc making sure they remain wet for the di...

Страница 15: ...lways start with a letter The model number and serial number are located on the underside of the seat section Model Number Serial Number 16796025 PXEXFT311873ACTTVN Ser PXR563146 PX FT 30 x 22 38 x 72...

Страница 16: ...amp 50 Hz Current Draw 2 7 amps 1 5 amps 3 3 amps Maximum Momentary Current Consumption 9 0 amps 5 0 amps 11 0 amps Voltage Output to Actuators 24 VDC 24 VDC 24 VDC Electric Shock Protection Class 1...

Страница 17: ...c environment in which radiated RF disturbances are controlled the customer or the user of the bed can help prevent electromagnetic interference by maintaining a minimum distance between portable and...

Страница 18: ...al mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical com mercial or hospital environ ment Voltage dips short inter ruptions and voltage vari...

Страница 19: ...RF transmitters as deter mined by an electromagnetic site survey ashould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the...

Страница 20: ...reedom PA 17349 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com made in the USA with US imported parts 3034177 CONFORMS TO ANSI AAMI STD ES60601 1 IEC STD 60601 1 3RD EDITION IEC STD 60601 1 2...

Отзывы: