- 4 -
3.3
Connectors
Symbol
Connector Name
Connector Type
COND
Socket for conductivity electrode
BNC
TEMP
Socket for temperature sensor
RCA
DC9V
DC9V power socket
Φ2.5 direct type
3.4
Stable reading display mode
When the measuring value is stable, smiley face icon appears on LCD, see
Figure-3. Please do not get the reading value or make calibration until the
icon appears and stays.
3.5
Data storage, recall, and deletion
3.5.1 Storage
When the measurement is stable, press key, the meter displays
M+
icon and storage numberings on LCD and stores the measuring data.
In Figure-4, the meter stores the first group of the measuring data. Press
again, to keep storing.
3.5.2 Recall stored data
In the measurement mode, long press key to recall the last stored
Measuring data. The meter displays
RM
icon and storage numberings.
In Figure-5, the meter is recalling the 28
th
stored measuring data.
Keep pressing and to recall successively the stored
measuring data. Press and hold and to quickly recall the stored
measuring data. Short press to go back to measurement mode.
3.5.3 Clear stored data
The meter can store 50 groups of data. Once the storage is full, if user
presses again, the
FUL
icon will be displayed on LCD (see Figure-6).
The storage needs to be cleared in parameter setting P4 to keep storing
data. The numberings in
M+
indicates the total number of stored data; The
Figure - 3
Figure - 5
Figure - 4
Figure - 6