- 4 - 

3.3 

Connectors

 

 

Symbol

 

Connector Name

 

Connector Type

 

COND

 

Socket for conductivity electrode

 

BNC

 

TEMP

 

  Socket for temperature sensor

 

RCA

 

DC9V

 

DC9V power socket

 

Φ2.5 direct type

 

 

 

3.4 

Stable reading display mode

 

When the measuring value is stable, smiley face icon          appears on LCD, see

 

Figure-3. Please do not get the reading value or make calibration until the 

 

icon appears and stays.

 

 

 

 

3.5 

Data storage, recall, and deletion

 

3.5.1  Storage 

When the measurement is stable, press            key, the meter displays

 

M+ 

icon and storage numberings on LCD and stores the measuring data.

 

In Figure-4, the meter stores the first group of the measuring data. Press

 

            again, to keep storing.   

         

 

 

 

3.5.2  Recall stored data 

In the measurement mode, long press            key to recall the last stored 

 

Measuring data. The meter displays 

RM

 icon and storage numberings. 

 

In Figure-5, the meter is recalling the 28

th

 stored measuring data. 

 

Keep pressing            and          to recall successively the stored 

 

measuring data. Press and hold          and        to quickly recall the stored 

 

measuring data. Short press              to go back to measurement mode.

 

 

3.5.3  Clear stored data 

The meter can store 50 groups of data. Once the storage is full, if user 

 

presses            again, the 

FUL

 icon will be displayed on LCD (see Figure-6). 

 

The storage needs to be cleared in parameter setting P4 to keep storing 

 

data. The numberings in 

M+ 

indicates the total number of stored data; The 

 

Figure - 3

 

 

Figure - 5

 

Figure - 4

 

Figure - 6

 

Содержание CON550

Страница 1: ...CON550 Benchtop Conductivity Meter Instruction Manual OAKTON INSTRUMENTS www 4oakton com v2 0...

Страница 2: ...e 4 3 5 Data storage recall and deletion 4 3 6 Max Min Function 5 4 Conductivity Measurement 5 4 1 Preparation 5 4 2 Conductivity Electrode Information 5 4 3 Information regarding Calibration 5 4 4 Co...

Страница 3: ...the meter s response time and measuring accuracy with stable reading display mode available The kit comes with a combination conductivity electrode with temperature sensor measuring conductivity and t...

Страница 4: ...dustproof and spillproof Size and Dimension 240 235 103 mm 1kg 3 Instrument Description 3 1 LCD Display 1 Measurement mode 2 Measurement value 3 Calibration icons and numberings 4 Measurement unit 5 T...

Страница 5: ...suring mode Cond TDS Long Press Enter parameter setting P1 P2 P6 Short Press Enter Calibration mode Cancel any operation and go back to measurement mode Short Press In calibration mode press to conduc...

Страница 6: ...st group of the measuring data Press again to keep storing 3 5 2 Recall stored data In the measurement mode long press key to recall the last stored Measuring data The meter displays RM icon and stora...

Страница 7: ...pter Plug the power adapter tightly into the DC9V socket The power s voltage should meet the requirements listed on the power adapter 4 2 Conductivity Electrode Information 4 2 1 Conductivity Electrod...

Страница 8: ...users can choose the calibration solutions that are close to the test samples to conduct 1 or 2 points calibrations 4 4 Conductivity Calibration take 1413 S as an example 4 4 1 Rinse the electrode in...

Страница 9: ...of the solution 4 6 2 During the process of calibration and measurement the meter has self diagnosis functions indicating the relative information as below Table 8 Table 8 Self diagnosis information o...

Страница 10: ...ting P1 Then press or to change menu from P1 to P2 P7 For details please see Table 10 Table 10 pH Parameter Setting P1 Select electrode s constant 1 0 10 0 0 1 1 Long press to enter P1 2 Press 1 0 fla...

Страница 11: ...lean Before taking a measurement rinse the electrode in pure water It is recommended to rinse it again in the sample solution When submerge the electrode in solution stir the solution briefly to elimi...

Страница 12: ...ENTS any malfunctioned or damaged product attributable to responsibility of OAKTON INSTRUMENTS for a period of three years from the delivery a six month limited warranty applies to probes This warrant...

Отзывы: